Maintenance

Storage

Consequences of using improper cleaning product

Cleaning products to avoid

5-4
7.
Dry thoroughly with a clean, dry, lint-free cloth and let air dry for at least 30
minutes. Do not apply heat.
NOTE
Drying times may vary based on the environmental conditions.
Remove batteries when the device is not in use, even for short periods of time.
Store in a dry well-ventilated area.
Hang the device using a holder if available.
If leadwires or cables are attached, hang them straight.
Do not coil leadwires or cables tightly around the device.
Appearance of waveforms when the device is not connected to a patient, causing
false alarms instead of a leads fail alarm and possibly not providing a visual or
audible leads fail alarm.
Brittle and breaking device case.
Overall system performance degradation.
Melting, dulling, or distorting the case.
Total medical device failure requiring replacement.
Unit malfunction.
Void warranty.
Cleaning products known to cause the types of problems listed above include, but are
not limited to:
®
Sani-Cloth
Wipes
®
Ascepti
Wipes
®
HB Quat
®
Clorox
Wipes (they do not contain bleach)
Over-the-counter detergents (e.g. Fantastic
Products that contain active ingredients and solutions similar to these products should
also be avoided.
NOTE
For additional information, refer to the How the Reach Us page in the manual for
contact information.
Patient Data Module
®
®
, Tilex
, etc.).
2030047-001A