Respiration

Temperature

2030047-001A
Respiration range limit:
Respiration rate accuracy
Impedance range:
Detection sensitivity range:
Respiration rate alarm
range:
No breath alarm range:
Number of channels:
Input specifications
Output specifications
Alarms:
Patient Data Module
Appendix A–Technical Specifications
1 to 200 breaths/minute
0 to 120 breaths/minute ± 1
121 to 200 breaths/minute ± 3
100 to 1500 Ohms at 52.6 KHz
0.4 to 10 Ohms impedance variation
1 to 200 breaths/minute
3 to 30 seconds
Up to 2 (with Y-adapter cable)
Probe type: Series 400 or 700 (determined by input cable)
Temperature range: 0° C to 45° C (32° F to 113° F)
Resolution: ±0.01° C (internal)
Parameters displayed: T1, T2
Error (independent of source):
±0.1° C with series 400 probes
±0.3° C with series 700 probes
User-selectable upper and lower limits
A-5