Dräger X-dock
Frequently Asked Questions
What is the service life of the internal datalogger of the X-dock station?
SOFTWARE
This highly depends on the configuration and the number of tests run.
With 2 x 2 GB (redundant) internal memory a system which e.g. carries out
40 tests per day and does not read out dataloggers or creates certificates,
can store local data for 10 years. If dataloggers are read out, more tests are
carried out or certificates created, this service life is certainly reduced.
How can I view the read out dataloggers of the gas detection instruments?
You can open the datalogger with the CC Vision which is free of charge
and thus convert it into a text file. The more comfortable way would be to
use Gas Vision. Via the X-dock Manager a direct link is already available
to open and view dataloggers with Gas Vision.
Do I still need CC Vision and Gas Vision?
For the time being CC Vision will remain the tool to configure the gas detection
instruments. However, from now on CC Vision is available free of charge.
For the time being Gas Vision will remain the tool to visualize the dataloggers
and to support you with the analysis.
For the time being means that in the medium term both software are to be
transferred to the X-dock Manager.
How do I have to configure a device in order to have the relevant test gas assigned?
CC-Vision may be used to set measured gas, calibration gas and bump test gas
if the device supports this functionality. X-dock then searches for and loads the
respective gas accordingly. Conversions occur automatically if the measured gas
does not match the calibration or bump test gas. If possible, measured gas is
recommended to use for all tasks.