Dräger X-dock
Frequently Asked Questions
Do I need a purge module for the exhaust gases (an additional pump to discharge exhaust gases)?
MISCELLANEOUS
No. The Dräger X-dock includes ONE output for exhaust gases. Here, a hose
with a length of up to 10m can be connected to discharge the exhaust gas.
A purge module is not necessary.
Can an adjustment be made when the bump test fails or the calibration interval is overdue?
Yes. When the option "auto repair" is activated, a failed gas application test is
"repaired" directly via an adjustment. If the calibration interval is overdue and this
option is activated, an adjustment is started directly and the gas application test
is skipped.
Why is there no X-am 7000 module?
The Dräger X-dock is specially attuned to the XXS sensors and supports
personal gas detection units such as X-am 125 and the Pac series which
are mainly used in larger device fleets. Unfortunately the Dräger X-am 7000
is not supported. For X-am 7000 we recommend using the Dräger Ecal.
Together with the Drägerware Workshop 3/5/7000 software thus also a joint
database for both systems can be operated.
Can I still use or purchase the bump test station?
Yes. The bump test station is still being supported and remains in our portfolio.
However, the results of these tests are not available in the database of the X-dock
system. The calibration date is certainly also updated after the next insertion in
the Dräger X-dock.