Storage Temperature
Operating
Temperature
Ingress Protection
Protocols
Connectors
Simultaneous
Connections
Venting
Barometric
Compensation
Barometer Accuracy
Alarms
Power
Wireless Setup
Logging Rate
Transmission Rate
Memory
Authorised distributor
In Australia:
For customer service, call 1300-735-292
To email an order, ordersau@thermofisher.com
To order online: thermofisher.com
1-970-498-1500
-20C to 60C
-20C to 50C (Li-MnO₂), 5C - 40C (Alkaline)
Device: IP68
System: Up to IP68 per antenna specification
Modbus over RS-485, SDI-12, Pulse low/high frequencies (max 40 khz)
1 In-Situ Twistlock (supports multiple instruments via Rugged Cable Splitter
or Load-bearing Universal Adaptor)
Up to 8 instruments
Built-in on all models, no desiccant required
Built-in on all models for automatic compensation of level readings
+/- 1 hPa
Configurable based on instrument readings and device parameters, second
reading/reporting schedule available when in alarm state
Max 75mA provided to connected instruments at 16V
Supported via Bluetooth Low Energy
1 minute to 7 days
5 minutes to 7 days
512 MB (on-board)