Press the kV+ switch to check all the codes stored. Maximum 20 codes can be
stored.

1.22.4 Program 4 : To Erase the errors that have occurred on the unit.

Select the program four in service mode and the display will read as below
Press the switch DL-serv on the D936 once.The display will read as below.
To erase the errors keep the mAs + switch pressed till the display reads
1.22.5 Program 5 : To set the display value at switch ON to Default Kv mAs
combination or last selected Kv mAs combination.
Select the program five in service mode and the display will read as below
Press the switch DL-serv on the D936 once.The display will read as below.
The default value can be set by selecting the desired kV, mAs combination.
Last selected combination can be stored by pressing the DL-serv switch twice.

1.22.6 Program 6 : To set the maximum kV and mAs value for the unit.

Select the program six in service mode and the display will read as below
Press the switch DL-serv on the D936 once.The display will read as below.
Set the required limit for kV and mAs using the kV /mAs increment and decrement
switches.

1.22.7 Program 7 : To set the max.inverter firing frequency(REG).

Select the program seven in service mode and the display will read as below
On activation of this program by pressing the DL-serv switch , the display indicates
Siemens Ltd. Med Division India
Pr
4
Era
Err
0
Err
Pr
5
las
Val
Pr
6
125
100
Pr
7
Adj
F
Service Instructions
Version 4.0
Service Manual
Page 12 of 48