1.22 Program Modes for servicing.

To put the unit into service mode, Short the ST link on the D915 and put ON the
unit.Use the Kv+ and Kv- switches to select the required program from 1 to 7.
To check the parameters in the program,press the DL-serv switch once.

1.22.1 Program 1 : To display the DC voltage on the capacitor.

Select the program one in service mode and the display will read as below
Press the switch DL-serv on the D936 once.
The display will read as below
XXX corresponds to the DC voltage on the capacitors.

1.22.2 Program 2 : To check the total number of exposures taken .

Select the program two in service mode and the display will read as below
Press the DL-serv switch and enter into the program.The display will be
XXX corresponds to the total number of exposures taken on the unit.
1.22.3
Program 3 : To check the type of codes that occurred in the unit
maximum 20 .
Select the program three in service mode and the display will read as below
Press the switch DL-serv on the D936 once.The display will read as below.
AA displays the serial no of the errors occurred in chronological order.
XX corresponds to the type of code.
Siemens Ltd. Med Division India
Pr
1
Cap
XXX
Pr
2
XXX
XXX
Pr
3
AA
XX
Service Instructions