4.7.5
– Rotate the hot tap fitting using the lock nut. (Upper stop of lock nut, with lettering 0 - CLOSE - ZU must be
visible; see detail A in Fig. 4-4).
– Disconnect transducer electrically according to instructions manual.
– Remove the nuts for the covers and carefully release the screws securing the transducer according to the fol-
lowing instructions.
DANGER
The mounting screws of the Sensyflow transducer may only be released when the hot tap fitting is in
!
disassembly position. Do not release the mounting screws when the fitting is in measuring position.
WARNING
Depending on design, a small amount of process gas (some cm
!
the transducer. For toxic or explosive gases, make sure that sufficient ventilation is provided.
WARNING
If when releasing the mounting screws you notice a large amount of gas escaping, the activities must
!
be stopped and screws retightened. Check whether the fitting is actually in disassembly position.
If it is positioned properly, disassembly of the transducer can only be performed after emptying and, if
necessary, cleaning the pipeline (defective hot tap fitting, see Maintenance).
– Remove the transducer from the hot tap fitting. Warning: Do not tip to the side.
4.7.6
After approx. 100 installation and disassembly procedures, the transducer O-ring gaskets for the PN 16 hot tap
fitting must be replaced. When working with dusty, abrasive or aggressive measured media, it may be necessary
to replace these more frequently.
Only ABB Service or qualified personnel of the operator is authorized to replace O-ring gaskets. For the latter
case, the warranty is no longer valid for ABB.
Fig. 4-11 shows the individual parts of the hot tap fitting:
Guide
tube (2)
Pipe
compo-
nent (1)
Fig. 4-11 Components of the integrated hot tap fitting
22
Thermal Mass Flowmeter FMT400-VTS, FMT400-VTCS (Sensyflow VT-S/VT-CS)

Disassembling the transducer with system in operation

Maintenance

Mounting and installation
3
) may be released when disassembling
Covers (3)
Trans-
ducer (4)
42/14-36 EN