b.
Amplitude Reading at 1701 MHz:
Frequency
Measured Values (dBm)
26 MHz -
28 MHz
c.
Amplitude Reading at 2145 MHz:
Frequency
Measured Values (dBm)
470 MHz -
472 MHz

14.8. Spectrum Analyzer Resolution Bandwidth Accuracy Test

BW
Span
VBW
Setting
3
6
Auto
MHz
MHz
1
2
Auto
MHz
MHz
300
600
Auto
KHz
kHz
100
200
Auto
KHz
kHz
30
60
Auto
kHz
kHz
10
20
Auto
kHz
kHz
3
6
Auto
kHz
kHz
1
2
Auto
kHz
kHz
300
600
Auto
Hz
Hz
100
200
Auto
Hz
Hz
30
60
3
Hz
Hz
Hz
10
60
3
Hz
Hz
Hz
40
Lower
Limit
2.7
MHz
900
kHz
270
KHz
90
KHz
27
kHz
9
kHz
2.7
kHz
.9
kHz
270
Hz
90
Hz
27
Hz
.9
Hz
Calculated IRS (dBc)
Calculated IRS (dBc)
Measured Values
Specification
(dBc)
–50
Specification
(dBc)
–60
Upper
Limit
3.3 MHz
1.1 MHz
330 KHz
110 KHz
33 kHz
11 kHz
3.3 kHz
1.1 kHz
330 Hz
110 Hz
33 Hz
11 Hz
Spectrum Master MM