Required Equipment - Fluke ProcessMeter 789 Calibration Manual

Hide thumbs Also See for ProcessMeter 789:
Table of Contents

Advertisement

General Specifications
Maximum voltage applied between
any jack and earth ground ..................... 1000 V
Battery Type ............................................ 1.5 V, 0-15 mA, AA, Alkaline
Storage temperature .............................. -40 °C to 60 °C
Operating temperature ........................... -20 °C to 55 °C
Operating altitude ................................... 2000 meters maximum
Frequency Overload Protection ............ 10
Temperature coefficient ......................... 0.05 x specified accuracy per °C for
Relative humidity .................................... 95 % up to 30 °C, 75 % up to 40 °C, 45 % up to
Vibration .................................................. Random 2g, 5 to 500 Hz
Shock ....................................................... 1 meter drop test
Power requirements ............................... Four AA batteries (alkaline recommended)
Size ........................................................... 10.0 cm X 20.3 cm X 5.0 cm (3.94 in X 8.00 in
Weight ...................................................... 610 g (1.6 lb)
Safety ....................................................... IEC 61010-1: 600 V CAT IV / 1000 V CAT III,
Electromagnetic Environment .............. IEC 61326-1: Portable
Electromagnetic Compatibility .............. Accuracy for all ProcessMeter functions is not
Applies to use in Korea only ................. Class A Equipment (Industrial Broadcasting &
[1] .. This product meets requirements for industrial (Class A) electromagnetic wave equipment
and the seller or user should take notice of it. This equipment is intended for use in
business environments and not to be used in homes.

Required Equipment

Equipment and software required to perform the procedures in this manual are identified
in Table 2.
If the recommended equipment model is not available, in some cases other equipment can
be substituted as long as it meets the specifications indicated.
To avoid safety hazards and equipment damage during the calibration
procedure, use the specified calibration equipment listed in Table 2.
Using unspecified equipment can jeopardize the calibration
verification test and pose safety hazards.
Unless otherwise indicated, all connection diagrams for the calibration
verification tests in this manual showing a calibrator or digital multimeter
use a Fluke 5500A calibrator or Agilent 3458A.
If you are using a different calibrator or DMM, make the connections
appropriate for that instrument.
6
V Hz max
temperatures <18 °C or >28 °C
0.1 x specified accuracy per °C for
temperatures <18 °C or >28 °C
50 °C, and 35 % up to 55 °C
X 1.97 in)
Pollution Degree 2
specified in RF field >3V/m
Communication Equipment
 Warning
Note
ProcessMeter™
Required Equipment
[1]
7

Advertisement