Current Measurement - Fluke ProcessMeter 789 Calibration Manual

Hide thumbs Also See for ProcessMeter 789:
Table of Contents

Advertisement

789
Calibration Manual

Current Measurement

1. Put the calibrator in Standby (
2. Put the UUT rotary switch in the
3. Connect the calibrator to the
4. Apply the values from the calibrator shown in Table 5 and compare the readings on
5. Connect the calibrator to the
6. Apply the values from the calibrator shown in Table 5 and Table 6 and compare the
7. Press J (BLUE) on the UUT to toggle to ac amps.
8. Apply the values from the calibrator shown in Table 7 and compare the readings on
14
Figure 4.
the UUT to the acceptable readings shown.
readings on the UUT to the acceptable readings shown.
the UUT to the acceptable readings shown.
UUT
Figure 4. Current Measurement Test Connections
) mode.
STBY
W
position
.
and
terminals on the UUT as shown in
COM
mA
and
terminals on the UUT.
COM
A
5500A
adm003F.EPS