Setup; Required Equipment - Fluke 1773 Calibration Manual

3 phase power quality analyzer
Hide thumbs Also See for 1773:
Table of Contents

Advertisement

Setup

Before you start the verification procedures or make calibration adjustments, refer to this
section for the equipment, system, and setup requirements.

Required Equipment

See
Table 2
for a list of requirements for the verification tests and calibration adjustment of the
Product. If the specific equipment is not available, use equivalent equipment that has the same
or better performance.
Item
Calibrator
Digital Multimeter (DMM)
Cable Assembly
Test Lead
17xx AUX Input
Calibration Cable
[1]
17xx Calibration Cables:
Voltage-to-Current Input
Cable Assembly
[1]
17xx Verification Box
USB Cable
Coil
Coil
Banana-to-Pin Adapter
[1] The 173x/174x calibration cables and verification box are not available from Fluke. See Equipment Assembly
for information on how to make these items. The 17xx calibration equipment is identical with the previous
173x/174x equipment.
Table 2. Required Equipment
Model
5522A with 8588A
Reference DMM
8588A or 8846A
3PHVL-1730
Pomona 6358
NA
NA
NA
[1]
USB A-to-USB C
5500A/COIL
Optional: 52120A
with Coils
NA
Pomona
Electronics 4690
3 Phase Power Quality Analyzer
Comments
5730A Calibrator, or
equivalent, if 8588A is
not available.
Voltage verification/
calibration and AUX
Adapter verification
Voltage Test Lead
3-Phase+N
For voltage PE
connection
for iFlex verification
5 turns
for AUX Adapter
verification
Setup
Used for:
Verification Adjustment
5

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

17751777