C Perfusion Index
Range: 0%~20%
D Resolution
SpO
: 1%, Pulse rate: 1bpm; Perfusion Index:0.1%
2
E Measurement Performance in Weak Filling Condition:
SpO
and pulse rate can be shown correctly when pulse-filling ratio is
2
0.4%. SpO
error is ±4%; pulse rate error is ±2 bpm or ±2% (select larger).
2
F Resistance to Surrounding Light:
The deviation between the value measured in the condition of man-made
light or indoor natural light and that of darkroom is less than ±1%.
G Power Supply Requirement: 3.6 V DC~4.2V DC.
H Optical Sensor
Red light (wavelength is 660nm, 6.65mW)
Infrared (wavelength is 880nm, 6.75mW)
I Adjustable Alarm Range:
SpO
: 0%~100%
2
Pulse Rate: 0bpm~254bpm

5 Installation

5.1 View of the front panel

8
Pulse sound
Power indica
Pulse rate va
Perfusion Ind
Menu button
Power butto
Pulse wavef