4-6.

Pulse

▶ It is for quick and short spin down.
Action
If you touch [Pulse] button and release at the point you want to
1.
stop, the centrifuge decelerates immediately
4-7.

Start/Stop

Action
4-7-1.
Start
1.
After setting RPM/RCF, Time and Temp., touch [Start] button.
During running, a 'Start LED' is turned on.
The instrument is running only when the door is closed.
4-7-2.
Stop
1.
In case of touching the [Stop] button, the operation is stopped.
4-8.

Emergency Door Open

For emergency door opening, you can use the Emergency Door Open Tool only
when the instrument is completely stopped.
The door can be unlocked manually with the Emergency Door Open Tool through the emergency door opening hole.
1.
Find the emergency door open hole in left side of the instrument.
2.
Insert Emergency Door Open Tool into the hole and push it until the door is released.
Manual opening should be performed only when spinning is completely stopped. Otherwise,
harmful damage will be accompanied to not only operators but samples.
After opening the door manually, it is recommended to wait until normal electricity comes back.