2. Product description

2.1 System components

Designation
ACCULAN dermatome
Cover
Sterile funnel
Rechargeable battery
Storage tray

2.2 Necessary components for the use

Designation
ACCULAN dermatome
Cover
Sterile funnel
Dermatome blades
ACCULAN battery charger
Rechargeable battery
Changing funnel

2.3 Intended use

The ACCULAN dermatome is used in dermatology/traumatology for
obtaining skin grafts.

2.4 Mode of operation

In the ACCULAN dermatome, the rotational speed of the motor is con-
verted, through an eccentric-tappet drive, directly and with minimum
loss into back-and-forth movement of the dermatome blade.

3. Preparation

Non-compliance with these rules will result in complete exclusion of
Aesculap liability.
Prior to use, inspect the ACCULAN dermatome and its accessories
for visible damage.
Use ACCULAN dermatomes and accessories only if they are in per-
fect condition.
10
Art. no.
GA 643
GA 624
GA 625
GA 646
GB 509 R
Art. no.
GA 643
GA 624
GA 625
GB 228 R
GA 627 M
GA 646
GA 644
ACCULAN dermatome GA 643

4. Working with the device

4.1 System set-up

Risk of infection and contamination! The device is
delivered in an unsterile condition.
Do not sterilize the rechargeable battery.
WARNING
Sterilize the device before use.
Inserting the rechargeable battery
7
Fig. 1
Hold the ACCULAN dermatome 1 in a way that the battery com-
partment is at the top.
Set the sterile funnel 6 in place.
Have the unsterile rechargeable battery 7 inserted by another per-
son, through the sterile funnel 6 into the battery compartment
(note the coding).
Remove the sterile funnel 6 and put it down in an unsterile room.
Insert the sterile cover 5, making certain that it clicks into place.
6