General Specifications - Fluke 150 Series Calibration Manual

Insulation testers
Hide thumbs Also See for 150 Series:
Table of Contents

Advertisement

General Specifications

Maximum Voltage Applied to any Terminal ... 600 V ac rms or dc
Storage Temperature........................................ -40 °C to 60 °C (-40 °F to 140 °F)
Operating Temperature .................................... -20 °C to 55 °C (-4 °F to 131 °F)
Temperature Coefficient .................................. 0.05 x (specified accuracy) per °C for temperatures < 18 °C or > 28 °C
Relative Humidity.............................................. Noncondensing
Vibration ............................................................ Random, 2 g, 5-500 Hz per MIL-PRF-28800F, Class 2 instrument
Shock ................................................................. 1 meter drop per IEC 61010-1 2nd Edition (1 meter drop test, six sides,
Electromagnetic Compatibility ........................ In an RF field of 3 V/M, accuracy = specified accuracy (EN 61326-
Safety ................................................................. Complies with ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA-C22.2
Certifications ..................................................... CSA per standard CSA/CAN C22.2 No. 61010.1-04; TUV per standard
Batteries ............................................................ Four AA batteries (NEDA 15A or IEC LR6)
Battery Life ........................................................ Insulation test use: Meter can perform at least 1000 insulation tests
Size..................................................................... 5.0 cm H x 10.0 cm W x 20.3 cm L (1.97 in H x 3.94 in W x 8.00 in L)
Weight ................................................................ 550 g (1.2 lb.)
IP Rating ............................................................ IP40
Altitude............................................................... Operating: 2000 m CAT IV 600 V, 3000 m CAT III 600 V
Over-Range Capability ..................................... 110% of range
Compliance to EN 61557 .................................. IEC61557-1, IEC61557-2, IEC61557-4, IEC61557-10
Model 1503 Accessories .................................. TL224 Leads
Model 1507 Accessories .................................. TL224 Leads
Model 1508 Accessories .................................. Lead Set PN 666602
Downloaded from
Elcodis.com
electronic components distributor
(< 64 °F or > 82 °F)
0 % to 95 % @ 10 °C to 30 °C (50 °F to 86 °F)
0 % to 75 % @ 30 °C to 40 °C (86 °F to 104 °F)
0 % to 40 % @ 40 °C to 55 °C (104 °F to 131 °F)
oak floor)
1:1997).
NO. 61010-1-04, and IEC/EN 61010-1 2nd Edition for measurement
category IV 600 V (CAT IV)
IEC/EN 61010-1 2nd Edition
with fresh alkaline batteries at room temperature. These are standard
tests of 1000 V into 1 MΩ with a duty cycle of 5 seconds on and 25
seconds off. Resistance Measurements: Meter can perform at least
2500 earth-bond resistance measurements with fresh alkaline batteries
at room temperature. These are standard tests of 1 Ω with a duty cycle
of 5 seconds on and 25 seconds off.
Non Operating (Storage): 12,000 m
TP74 Probes
Clips PN 1958654 (red) and PN 1958646 (black)
Holster
Remote Probe
TP74 Probes
Clips PN 1958654 (red) and PN 1958646 (black)
Holster
Remote Probe
Clips PN 1670641 (red) and PN 1670652 (black)
Holster
Remote Probe
Insulation Testers
General Specifications
3

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

150315071508

Table of Contents