HP 8340A Service Manual page 21

Synthesized sweeper 10 mhz to 26.5 ghz
Hide thumbs Also See for 8340A:
Table of Contents

Advertisement

SERIAL NUMBER; 9999
================================================
1. DWELL TIME
2. YTM BX DLY 2
3. YTM BX DLY 3
4. YTM BX DLY 4
5. YTM DLY I
6. YTM DLY 2
7. YTM DLY 3
8. YTM DLY 4
9. YTM GAIN 1
10. YTM GAIN 2
11. YTM GAIN 3
12. YTM GAIN 4
13. LO SLOPE
14. HI SLOPE
15. 9GZ SLOPE
16. 20GZ SLOPE
17. ATIO OFFSET
18. AT20 OFFSET
19. AT30 OFFSET
20. AT40 OFFSET
21. AT50 OFFSET
22. AT60 OFFSET
23. AT70 OFFSET
24. AT80 OFFSET
25. AT90 OFFSET
26. ATIO SLOPE
27. AT20 SLOPE
28. A T30 SLOPE
29. AT40 SLOPE
30. AT50 SLOPE
31. AT60 SLOPE
32. A T70 SLOPE
33. AT80 SLOPE
================================================
================================================
Procedure for manually entering calibration data into the HP 8340A:
1.
Push the following sequence of KEYS:
Instrument Preset SHIFT MHz 1 2 Hz SHIFT KHz 2 2 Hz
2.
Note the Entry Display will indicate the Calibration Constant number and value.
3.
Enter via the KEY BOARD or DATA KNOB the correct value for the fi r st Calibration Constant indicated in the
display.
« Terminate KEYBOARD entries with the Hz key »
4.
Go to the next Calibration Constant by pushing the UP step key. The next constant can then be entered. Do not
enter the "CHECKSUM" Constant. (This is computed automatically)
5.
The Step Keys can be used to move from one Calibration Constant to another to either check them or to correct
them.
6.
After all entries have been made, check that all numbers are correct by using the step keys to review and verify
them.
7.
Allow instrument to warm up for 1/2 hour and make sure that nothing is connected to the Stop Sweep connector
on the rear panel. Push SHIFT PEAK to perform an Automatic Tracking Calibration. This step may modify
the "YTM GAIN n" & "YTM BX GAIN n" consts.
8.
The CALIBRATION data should be permanently stored in the Non Volatile Protected Memory by pushing the
following key sequence: SHIFT MHz 14 Hz SHIFT KHz 5349 Hz PRESET.
Figure
8A-4.
8-20
Model 8340A - Service
8340A CALIBRATION CONSTANTS
50
34. A T90 SLOPE
112
35. unused
131
36. unused
37. unused
125
89
38. unused
100
39. ADC GAIN LO
117
40. ADC GAIN HI
104
41. MAX SWEEP RATE
755
42. ADC OFFSET
859
43. AM OFFSET
772
44. LVL DAC OFF; INT
990
45. L VL DAC OFF; EXT
40
46. LVL DAC GAIN; LO
119
47. LVL DAC GAIN; HI
100
48. LVL DAC GAIN; EXT
215
49. PWR SWP GAIN
-2
50. YTM BX GAIN 1
-5
51. YTM BX GAIN 2
-1
52. YTM BX GAIN 3
-4
53. YTM BX GAIN 4
-7
54. STOP LIMIT
-2
55. START LIMIT
-5
56. PRESET OPTION
-7
57. HPIB ADDRESS
-10
58. RETRACE DWELL
-2
59. ATTEN CONFIG
-4
60. CONFIGURATION
-3
61. SERIAL#
-5
62. ATIO SLP 20GZ
-7
63. A T20 SLP 20GZ
-60
64. A T30 SLP 20GZ
-8
65. AT40 SLP 20GZ
-10
66. A T50 SLP 20GZ
Sample Calibration Constants Hard Copy (Found Inside Instrument)
Scans by HB9HCA and HB9FSX
8340AREV 08 JAN 85
-12
67. AT60 SLP 20GZ
II
0
II
68. A T70 SLP 20GZ
0
69. AT80 SLP 20GZ
II
70. A T90 SLP 20GZ
0
II
II
71. YTM OFFSET I
0
-33
72. YTM OFFSET 2
73. YTM OFFSET 3
-28
74. YTM OFFSET 4
600
-36
75. YTM TC GAIN
0
76. YTM TC OFFSET
33
77. YTM TC BKP GAIN
2
78. YTM TC BKP FREQ
8
79. OI MODEL
3
80. unused
26
81. LOCKWORD
I
82. LOCKWORD ENABLE
890
83. unused
791
84. unused
1030
85. unused
990
86. unused
20000
87. unused
10
88. unused
0
89. unused
531
90. unused
0
91. unused
20980
92. unused
23
93. unused
9999
94. unused
4
95. unused
5
96. unused
5
97. unused
8
98. MODEL#
9
99. CHECKSUM
10
13
13
18
1024
1024
1024
1024
25
1000
175
50
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
9557