Progetti EPG View/Par. C./ Int. User Manual page 8

1-3 channels electrocardiograph with graphical display and printer
Table of Contents

Advertisement

Position the electrodes according to the following scheme:
C1: on the 4
C2: on the 4
C3: on the 5
C4: on the 5
C5: on the left anterior axillary, at the same level as C4.
C6: on the left mid-axillary at the level of C4.
Peripheral electrodes:
N: Right leg
R: Right hand
F: Left leg
L: Left hand
ECG DERIVATIONES ACCORDING TO IEC and AHA:
IEC
----------------------------------------------------------------------------------------------------------
N Black
R Red
C1 White/Red
C2 White/Yellow
C3 White/Green
C4 White/Brown
C5 White/Black
C6 White/Violet
L Yellow
F Green
It is important not to use electrodes of different material as this can create polarisation
currents and disturb the base line of the ECG signals.
3.10 IMAGE FREEZING
Pushing the key "Freeze"
will continue for 10 seconds (RECORD).
When the acquisition is finished (MEMORY), pushing the
possible to review the traces memorized and to print them pressing the key .
PROGETTI S.r.l. - Via B. Buozzi, 28 10024 Moncalieri (To) Italy – www.progettimedical.com - info@progettimedical.com
th
intercostal space on the right parasternal part.
th
intercostal space on the left parasternal part.
th
rib, between C2 and C4.
th
intercostal space, on the left hemiclavicular region.
it is possible to stop the display. The data acquisition
EPG View USER MANUAL V. 03-10 Pag. 8
AHA
RL Green
RA White
V1 Brown/Red
V2 Brown/Yellow
V3 Brown/Green
V4 Brown/Blue
V5 Brown/Orange
V6 Brown/Violet
LA Black
LL Red
and
keys it is

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents