Troubleshooting; Causes And Remedies - Progetti EPG View/Par. C./ Int. User Manual

1-3 channels electrocardiograph with graphical display and printer
Table of Contents

Advertisement

6. TROUBLE SHOOTING
TROUBLES
Pressing the "Freeze" key the graphic does
not stop
Graphic not displayed
Faded hardcopy
No hardcopy
Batteries not recharging
Device does not turn on
Faulty paper flow
Baseline drift
Muscle tremors
Interferences on trace
PROGETTI S.r.l. - Via B. Buozzi, 28 10024 Moncalieri (To) Italy – www.progettimedical.com - info@progettimedical.com
EPG View USER MANUAL V. 03-10 Pag. 11

CAUSES AND REMEDIES

After pressing the "Freeze" key wait the
cursor to reach the end of display, then press
the other keys.
Batteries discharged
Non original paper or batteries discharged
Wrongly positioned paper roll, thermosensitive
side up.
Verify power cable connection
Recharge the batteries
End of paper roll, non original paper or
wrongly positioned paper roll
Dirty electrodes surface, insufficent skin
contact or patient movement
Patient not relaxed; filter non active
Faulty patient cable; dirty electrodes surface
or insufficent skin contact