GE ApexPro Operator's Manual page 62

Telemetry system
Hide thumbs Also See for ApexPro:
Table of Contents

Advertisement

System setup
4-12
Heart rate alarm limits (high/low):
Adult—150/50
0–2 years—200/90
3–10 years—180/60
11–13 years—150/50
ST measurement:
Adult— J+ 60ms
0–2 years— J+ 30ms
3–10 years— J+ 40ms
11–13 years— J+ 50ms
PVC limit is 6
1X size
Pace off
Arrhythmia On (arrhythmia changes with age)
ST off
Graph leads II and V
25 millimeters per second speed (fixed)
Display aspect ratio of 0.4 (fixed)
Alarm Graph location at the CIC Pro center where patient was admitted
TTX (manual) graph location at the CIC Pro center where patient was admitted
Print window location at the CIC Pro center where patient was admitted
Telemetry alarm control defaults
This option sets the telemetry default alarm limits and alarm level settings. In user
mode, all of the controls are view-only. You must be in Service mode to set the
Telemetry Alarm Control Defaults at the CIC Pro center.
To view the telemetry unit default settings, click CIC Setup > Telemetry Alarm
Control Defaults.
ApexPro™
2001989-200E