Single-Lead ECG telemetry data

2001989-200E
Single-Lead ECG may be acquired using a 3-, 5-, or 6-leadwire set. However, only a
Single-Lead ECG is transmitted or processed.
Single-Lead analysis is beneficial when troubleshooting pacemaker detection and/or
arrhythmia detection. Single-Lead analysis must always be used when monitoring
with a 3-leadwire set. Single-Lead analysis can be set up as a unit default. Refer to
Customizing the system on page 4-4
NOTE
When acquiring Single-Lead ECG data using a 5- or 6-leadwire set, it is not
necessary to connect the V leads or the right leg lead to the transmitter or to the
patient.
The following constraints apply when using Single-Lead ECG telemetry data.
Function
change the displayed
lead
Lead Analysis
select a V lead
select displayed leads
from a single viewer
select graph waveforms
ApexPro™
Viewing real-time patient data
for more information.
Single-Lead Constraints
The factory default Display Lead is lead II.
Contact your local service representative to change the
default displayed lead.
Display Lead appears to be selectable at the CIC Pro
center. However, your selection is temporary and will
revert back to the transmitter's default displayed lead.
NOTE
When the clinical situation dictates monitoring a lead
other than the default lead, you can move the leads
and/or electrodes to view a different lead. Be aware
that the label on the display and on the printout will
show the default lead label.
Multi-Lead analysis may appear to be selectable at the
CIC Pro center. However, your selection is temporary and
will revert back to the Single-Lead analysis mode.
V leads may appear to be selectable at the CIC Pro
center. However, your selection does not change the
transmitter's default displayed lead.
Leads other than the default displayed lead may appear to
be selectable at the CIC Pro center. However, your
selection is temporary and will revert back to the
transmitter's default displayed lead.
Leads other than the default displayed lead may appear to
be selectable at the CIC Pro center. However, you must
select the transmitter's default displayed lead to obtain a
graph of the waveform.
7-17