Operating Conditions

STATUS INDICATORS

While the PROCESSOR is operating, the condition of the PROCESSOR is communicated to the
operator as outlined below:
• When the READY INDICATOR illuminates and neither the WAIT INDICATOR nor SERVICE
INDICATOR is illuminated or blinking, it indicates that:
– The developer and the DRYER are at the correct temperature.
– You may feed film.
• When the READY INDICATOR illuminates and the WAIT INDICATOR is blinking 3 times, it
indicates that:
– The solutions are at the correct temperature.
– The DRYER has not reached the set-point temperature.
– Emergency films may be run, but may exit wet or damp.
• When the WAIT INDICATOR illuminates and remains illuminated while the PROCESSOR is
operating, it indicates that:
– A film is being fed and is not yet off the FEED TRAY.
– The FILM SENSORS may be blocked by dirt or other debris, or may be defective.
• When the WAIT INDICATOR blinks on and off while the PROCESSOR is operating, it indicates
that:
– There is a temporary error or warning condition. See page .
– There is an error condition that the operator may be able to correct. See Table on pages
through .
– The pattern of "ON" and "OFF" flashes indicates the wait error code.
• When the SERVICE INDICATOR blinks on and off while the PROCESSOR is operating, it
indicates that:
– There is an error condition that the operator may be able to correct. See Table on pages to .
– The pattern of "ON" and "OFF" flashes indicates the service error code number.
• When the SERVICE INDICATOR illuminates and remains illuminated while the PROCESSOR
is operating, it indicates that:
– The PROCESSOR has been left in Diagnostic Mode.
– To exit the Diagnostic Mode, open the ELECTRICAL BOX and move all the DIAGNOSTIC
SWITCHES on the 100 CIRCUIT BOARD to the "OFF" position.
The STATUS INDICATORS help you to diagnose problems with the PROCESSOR. Check them
before servicing the PROCESSOR. See the tables on the pages that follow.
981093 – October 1994
Operating Conditions