Normal Operation - Kodak X-Omat M43 Diagnostics Manual

Processors
Hide thumbs Also See for X-Omat M43:
Table of Contents

Advertisement

SITE SPECIFICATIONS
ERROR CONDITION Blink Patterns
Tables and show the blink patterns of the STATUS INDICATORS. Use these codes to diagnose problems before
servicing the PROCESSOR.
Warning
• Only qualified service providers should attempt the diagnostic procedures outlined in the following tables.
Doing the following procedures will expose you to high voltage components. Use caution when operating the
PROCESSOR with the COVERS removed. The BLOWER MOTOR and the MAIN DRIVE MOTOR energize
intermittently.
• In PROCESSORS having a serial number of less than 350 or without Mod 1 installed, the BLOWER
MOTOR and the MAIN DRIVE MOTOR will energize intermittently if all of the conditions below are met.
– if the TOP COVER is removed
– if the INTERLOCK TOOL TL-4740 is inserted into the AC INTERLOCK SWITCH S5
• In PROCESSORS having a serial number of 350 or higher or PROCESSORS with Mod 1 installed, the
BLOWER MOTOR and the MAIN DRIVE MOTOR will energize intermittently if all of the conditions below
are met.
– if the TOP COVER is removed
– if the INTERLOCK TOOL TL-4740 is inserted into the AC INTERLOCK SWITCH S5
– if a MAGNET is placed on DC INTERLOCK SWITCH S6
Important
The DC INTERLOCK SWITCH S6 is only found in PROCESSORS that either have a serial number of 350 or
higher, or have Mod 1 installed.
Error Condition - WAIT INDICATOR Blink Patterns
Error Condition
PROCESSOR ready
Ready - On
Wait - Off
Service - Off
No error, film being fed
Ready - On
Wait - On
Service - Off
18
Corrective Action
NORMAL
OPERATION
NORMAL
OPERATION
October 1994 – 981093