Troubleshooting - Satelec Acteon Mini LED III Supercharged User Manual

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
MLED III Supercharged U2
Before cleaning the handpiece, insert the protective
cap in the place of the light guide, supplied to ensure
that liquid does not enter the handpiece.
Avoid using cleaning and disinfection products
containing flammable agents (or other corrosive
agents such as acetone, chlorine or bleach).
Otherwise, ensure that the product completely
evaporates and that there are no combustibles on the
unit and its accessories before operation.
Do not use abrasive products to clean the unit.
Do not immerse the unit.
Never use ultrasound cleaning, whether for the Mini
LED SuperCharged or its accessories.
None of the accessories are delivered in a sterilized
state.
Only the light guide and the flexible protective shield
are sterilizable.
Before sterilization, check the cleanliness of your
autoclave and the quality of the water used.
After each sterilization cycle, immediately remove
the items from the autoclave to reduce the risk of
corrosion of metallic parts.
It is necessary to leave the sterilized items to cool
down to the ambient temperature and dry before re-
using them.
It is recommended to sterilize the items identified as
sterilizable by Satelec individually in the sterilization
bags foreseen for this purpose.
To maintain the sterile and aseptic state of
accessories, make sure that they are kept in
hermetically-sealed bags or containers suited to use
in dentistry.
The maintenance and/or sterilization instructions
that follow must be applied before each use of the
unit.
8.1 – Pre-disinfection/cleaning
Clean and disinfect the body, light guide, flexible shield
and rigid protection shield, and the charging base of the
Mini
LED
SuperCharged
cleaning/disinfecting wipes based on alcohol,
amphoteric disinfectant and biguanide (refer to the
manufacturer's instructions) for at least two minutes.
Leave the product to act for at least 15 minutes.
Use wipes with CE marking, or compliant with any
standard that may be required by national regulations.
8.2 - Drying
Dry using a clean, single-use, non-woven cloth to
remove any liquid traces.
V5:Mise en page 1
22/02/2010
8.3 - Packing
Pack in single-use sterilization bags or sleeves compliant
with the requirements defined in the EN ISO 11607-1
standard, or compliant with any standard that may be
required by national regulations.
8.4 - Sterilization
The optic fiber and the flexible protective shield for the
Mini LED SuperCharged must be sterilized individually in
an autoclave according to the following parameters:
- Autoclave: Type B compliant with the EN 13060
- Sterilization temperature: 134°C.
- Sterilization steady state: 18 minutes.
- Pressure: 2 bar minimum.
CAUTION: Not all autoclaves can reach 134°C. Not all
autoclaves perform a pre-depressurization. For further
information on the applicable sterilization instructions,
consult the autoclave manufacturer.
8.5 - Storage
After this, store the sterilized items in a dry, dust-free
place.
Before re-use, in the event of nonconforming packaging
integrity, re-package and re-sterilize according to the defined
protocol.
IX - TROUBLESHOOTING
In the event of a problem, before contacting the after-
sales service of SATELEC or the supplier:
- Ensure that the base is correctly connected to the AC
- Check that at least one of the four round battery
using
ready-to-use
- In the event of a faulty battery (see VII – INTERLOCKS)
- Intensive use of the lamp can result in high
- After each use, check that there is no composite
12
12:47
Page 12
standard.
supply to ensure that the battery is charged normally.
If the Mini LED SuperCharged s holder is not lit despite
being correctly connected, contact the after-sales
service of SATELEC or the supplier.
charge level indicators is visible on the LCD screen
before pressing the ON/OFF button.
an automatic interlock system will prevent the unit
from operating. If this is the case, recharge the
battery by placing the lamp on its holder or by using a
second fully-charged battery.
temperatures inside the unit. If this happens, an
automatic interlock system prevents the unit from
operating (see VII – INTERLOCKS). Leave the unit at
rest for a few minutes to allow it to cool down.

Advertisement

Table of Contents
loading

Table of Contents