Summary of Contents for Satelec Acteon Mini LED ScanWave

  • Page 1 ® User’s manual...
  • Page 2 igure 2 ● NG39EN020E ● 05/2013 ● (13) ● V5 ● J05451 ●User’s Manual ScanWave...
  • Page 3: Table Of Contents

    ENGLISH I – DESCRIPTION ..........4 II – ROUTINE OPERATION ......... 6 III – CONFIGURATION ........7 IV – SECURITIES ..........7 V - ACCESSORIES ........... 8 VI - TROUBLESHOOTING GUIDE ......8 VII – SYMBOLS AND ABBREVIATIONS ....10 User’s Manual ScanWave ●...
  • Page 4: I - Description

    1.3 lcd This is the user’s manual for your Scanwave device. screen This document must be used together with the general The LCD screen is backlit and allows the display of the user’s manual for SATELEC curing lamps. different pieces of information required by the user. I –...
  • Page 5 screen displays the message «Lb» (Low battery), the 1.8 t echnical sPecifications handpiece beeps 4 times, and the start of a curing Model name ScanWave cycle will be inhibited. Manufacturer SATELEC Medical category according directive 93/42/CEE Handpiece Weight 185g Dimensions Ø34 x 220mm Operating cycle •...
  • Page 6: Routine Operation

    Minimum wavelength 650nm • To switch from one mode to the other, quickly press Maximum wavelength 660nm the «Mode» button located on the left under the LCD Intensity < 1mW screen. Class • To scan and select one of the cycles contained in Exposure time As long as user presses the these modes, quickly press the «Cycle»...
  • Page 7: Configuration

    an icon indicates the type of light emission selected: select the pen-style hold. Initiation is then only continuous, flashing or gradual. possible by pressing the ON/OFF button and the screen displays PE: «Pen-Style». • Place the optical guide as close as possible to the surface of the equipment to be photocured but avoid contact as this may reduce the quality of the curing process.
  • Page 8: Accessories

    It is then recommended that the user leaves the handpiece to cool for a few minutes, until the screen displays the selected time cycle again and then the light can be switched on again. V - ACCESSORIES The following accessories are available for ScanWave •...
  • Page 9 Symptom Possible causes Solutions Defective laser diode or defective Return to SATELEC after-sales department Laser defect buttons or no light power Light guide defective and / or Clean light guide and / or return to SATELEC after-sales dirty department AC power wall outlet defective Contact your electrician AC power adapter defective Return to SATELEC after-sales department...
  • Page 10: Symbols And Abbreviations

    VII – SYMBOLS AND ABBREVIATIONS SYMBOL DEFINITION Alternating current Direct current Do not discard as household waste Please refer to the accompanying documentation «ON»/«OFF» (pushbutton) Type B Class 2 laser equipment Protective glasses recommended CE marking No access for persons with active implant as pacemaker United States Federal Law restricts on its territory the use of this unit...
  • Page 11 User’s Manual ScanWave ● J05451 ● V5 ● (13) ● 05/2013 ● NG39EN020E ● 11...
  • Page 12 J05451 ● V5 ● (13) ● 05/2013 ● NG39EN020E 17 av. Gustave Eiffel • BP 30216 • 33708 MERIGNAC cedex • FRANCE Tel + 33 (0) 556 34 06 07 • Fax + 33 (0) 556 34 92 92 E-mail : satelec@acteongroup.com • www.acteongroup.com...