SERIES 5 SIDE DELIVERY SYSTEM
Instructions For Use

GENERAL INFORMATION

INTENDED APPLICATION AND USE
The DCI Edge dental delivery system is a device intended to support the instruments used by the dental practitioner, delivering those instruments
to an accessible position during a dental procedure. This device also controls and is the means of delivering compressed air, water, and low voltage
electricity to a variety of instruments commonly used in dental practice. The delivery system is classified as a Class 2A device under rule 11 of
Annex IX of the MDD 93/42/EEC. Accordingly, the provisions of Annex II apply.
PRODUCT IDENTIFICATION
This device can be identified by the label on the
underside of the unit head. This label states the
model number, serial number, electrical specifications,
manufacturing date, software version, and safety clas-
sifications. Note the sample label at right.
ELECTRICAL SPECIFICATIONS
100-240VAC
.6 Amps
50-60Hz
ACCESSORY DEVICES:
Power Optics:
Operation: Intermittent
Duty cycle: 20 sec ON, 10 sec OFF , 10X/hr
ELECTRICAL DEVICE INTERFERENCE
To guarantee the operational safety of electromedical devices, it is recommended that the
operation of mobile radio telephones in the medical practice or hospital be prohibited. Strong
EMI sources such as electro-surgery units or x-ray units may affect performance. If performance
problems occur, move the unit to another electrical circuit or physical location.
TECHNICAL DOCUMENTATION
The manufacturer will make available upon request circuit diagrams, component part lists,
descriptions, and calibration instructions to technical personnel responsible for the installation and
service of this equipment.
PREVENTATIVE INSPECTION
The performance of the equipment can be affected by use over time. Periodically inspect the water and air lines for visible cracks or cuts and inspect
for loose fittings and fasteners which could lead to leaks or other poor performance characteristics. Inspect joints and tensioning screws as a regular
maintenance item to ensure proper positioning of the device.
4
IEC Medical Classification 2A
TYPE: BF Power optics
Splash Protection: IPX0
AIR AND WATER SUPPLY
WATER
AIR
Air Quality: Dry and clean
Water Quality: Potable
Pressure: 80-105 psi
Hardness: 7.2 - 7.8 pH
Pressure: 40-80 psi
ENVIRONMENTAL CONDITIONS
Operating Conditions:
Temperature: 67-76° F
Temperature: -68 - 122° F
PRODUCT DISPOSAL
Contact your local, authorized dealer
for disposal of this device to ensure
compliance with your local
environmental regulations.
Humidity: 20-60%
Shipping conditions:
Humidity: 10-90%