MAINTENANCE

GENERAL CLEANING
Equipment can be cleaned with a solution of mild detergent and warm water. A variety of surface disinfectants are
available for use in dental treatment rooms. Some of these can cause discoloration of painted, plated or anodized
surfaces with repeated use. This can be minimized by careful adherence to the disinfectant manufacturer's instructions
and by frequent washing with soap and water.
HANDPIECE FLUSH - DAILY MAINTENANCE
The dental unit is equipped with a handpiece flush system that allows you to periodically flush fresh water through the
handpiece tubings. The need for this is caused by the low flow of water through the tubings during normal use, which
can lead to stagnation and the potential growth of "biofilm" contamination.
It is recommended that you flush the tubings at the beginning and end of each day. This may be done with or without
the handpieces installed, but having the handpieces on the tubings will restrict flow, so a longer flush time will be
required. All of the tubings are flushed simultaneously. Hold them together and direct them into a basin, sink or
cuspidor to catch the water. Flip and hold the flush toggle.
Allow adequate time for fresh water to make its way through the entire system and displace all standing water. The
American Dental Association and the Centers for Disease Control can provide additional recommendations regarding
this procedure, including information on frequency and duration of flushing and the use of antibacterial solutions in the
self-contained water system.
CAUTION: Do not use powdered cleansers, scouring pads or abrasive scrubbers on any of the
painted, plastic or metal surfaces of this dental unit. To remove dried-on material, use a soft bristled
brush and a solution of mild detergent.
WARNING: In order to mitigate the risk of cross-contamination between patients, disinfection
and sterilization of handpieces must occur after each dental procedure. Refer to the handpiece
manufacturer's instructions and recommendations for sterilization or disinfection procedure.
SERIES 5 SIDE DELIVERY SYSTEM
Instructions For Use
15