Resistor
Range
Resolution
600.0Ω
0.1Ω
6.000kΩ
0.001kΩ
60.00kΩ
0.01kΩ
600.0kΩ
0.1kΩ
6.000MΩ
0.001MΩ
40.00MΩ
0.01MΩ
There is a little rolling less than ±50 digits when
measuring > 10.00 MΩ.
Overload Protection: AC/DC 1000V
Continuity
Range
Resolution
600.0Ω
0.1Ω
Continuity: Built-in buzzer sounds when measured
resistance is less than 20Ω and sounds off when
measured resistance is more than 200Ω, Between 20Ω
to 200Ω the buzzer maybe sound or off either.
Continuity Indicator: 2.7kHz Tone Buzzer
Response Time of Buzzer: < 500μsec.
Overload Protection: AC/DC 1000V
Diode
Range
Resolution
1.500V
0.001V
Open Circuit Voltage: Approx. 1.8V
Overload Protection: AC/DC 1000V
23
Accuracy
±(1.0% +5D)
±(1.0% +2D)
±(1.5% +5D)
Accuracy
±(1.0% +5D)
Accuracy
±(1.0% +5D)