Labware considerations

Acceptable microplates
Lidded microplates
G5580A BenchCel Microplate Handler Quick Guide
The BenchCel device is designed to handle labware that comply with the standards
ANSI/SLAS 1-2004 (R2012) through ANSI/SLAS 4-2004 (R2012). For use of
nonconforming labware, contact Agilent Technical Support.
The BenchCel device uses gripping mechanisms to hold microplates securely and
repeatably in the labware rack and in the robot arms. The BenchCel device typically
holds the microplates halfway between the top of the microplate and the top of the
microplate skirt (5 to 10 mm above the bottom of the microplate).
In the following figure, notice the gripper-microplate contact point
Figure
Labware rack detail with microplate held by the stacker grippers (front view)
Microplates that do not have lids or have shallow lids (lids that do not reach the
microplate skirt) provide enough clearance to allow secure and repeatable gripping.
Microplates with deeper lids can be more challenging, because the microplate must be
held by the skirt. If the skirt is too flexible, the stacker grippers will bend the skirt. The
bent skirt can grip the microplate lid stacked beneath, inadvertently removing the lid.
Some labware vendors might offer alternative lids that are shallower. Contact the
vendor for details.
Figure
Lidded microplate examples
Item
Description
1
No lid: excellent gripper clearance
2
Shallow lid: good gripper clearance
3
Deep lid: no gripper clearance, must be held by skirt
1
2
3