7. INSTRUCTIONS FOR PLATFORM EQUALIZATION
1. Execute 20 clockwise turns at least of the screws of all the trimmers "T", in order to obtain the
maximum signal on all the cells.
EQUALIZATION BOARD
T
2. Perform the calibration with central mass (ex. full scale).
3. Set the indicator on ×10 visualization.
4. Position the sample weight (approximately
of the cell's maximum capacity) on all the angles
(1,2,3,4) alternatively, and identify the cell indicating the lowest weight.
DISPOSITION OF THE CELLS ON FPW PLATFORM
12