GENERAL SPECIFICATIONS
1.
MS8302E is a stable, safe, reliable compact digital handheld 6000 count
auto-ranging multimeter. This meter can measure AC/DC voltage, AC/DC
current, resistance, capacitance, frequency, duty cycle, temperature,
diodes and continuity. This meter is ideal for many situations, whether
you're a professional or causal user.
• Operating Altitude: 2000m
• Relative Humidity: 75% max operating
• Operating Temperature: 0°C~40°C/32°F~104°F (<80% RH)
• Storage Temperature: -10°C~60°C/14°F~140°F (<70% RH)
• Accuracy Temperature: -18°C~28°C/64°F~82°F (<80% RH)
• Temperature Coefficient: 0.1x(specified accuracy)/°C (<18°C or >28°C)
• Sampling Frequency: approx. 3 times/sec.
• Fuse Protection: µA/mA input: FF 600mA H 600V
10A input: FF 10A H 600V
• DC/AC Voltage: 600V
• DC/AC Current: 10A
• Resistance: 40MΩ
• Diodes: 3.2V DC
• Continuity: <50Ω
• Temperature: -20°C~1000°C/-4°F~1832°F
• Battary Voltage: 1.5V/9V
• LCD Display: 3 ¾ digit display (4000 counts)
• Product Supply: 9V battery
• Product Size: 150mm×74mm×48mm / 6.3"×2.9"×1.9"
• Product Weight: 482g / 1.06lb
• Safety Rating: CAT III 600V; pollution degree: II
• Safety Standards: IEC61010-1
• Pollution Degree: 2
• Accuracy: ±(of reading + # of least significant digits)
1
2.
To avoid electric shock and injury or damage to the meter, observe the
following safety methods:
Check the meter before use to make sure there was no damage during
•
transit.
•
Check that the insulation on the test leads is not damaged and/or wires
are not exposed.
•
If any faults or abnormalities are observed, the meter should not be used
and should be checked out prior to use.
•
Never exceed the protection limit values indicated in specifications for
each range of measurement.
•
Always be careful when working voltages above 60V DC or 30V AC rms,
keep fingers behind the probe barrier while measuring.
•
Make sure the rotary switch is in the correct position before measurement.
•
Never use the meter in an environment with explosive gas, vapor or dust.
•
Always keep fingers behind probe barriers when making measurements.
•
When connecting test leads to a circuit, connect the black test lead first,
then the red lead. Disconnect in the opposite order.
•
Turn off power and discharge all capacitors first before measuring
resistance, continuity or diodes.
•
Failure to follow safety guidelines could compromise the safety features
of this meter.
•
Do not use the meter without the battery cover in place.
•
Replace the batteries as soon as the low battery symbol "
false reading that could lead to electric shock and injury.
•
Remove test leads from all circuits before opening the battery cover.
WARNINGS
" to avoid
2