CAUTION
If all four casters are not locked, swivel and wheel rotation will allow bed motion with little
or no resistance. This can result in loss of balance, fall and personal injury. Ensure that all
four casters are locked to avoid any bed movement during resident transfer and/or use. The
central locking mechanism can be used for most situations where locking casters are needed.
However, movement near the head section of the bed may still result if force is applied. The
bed will be most secure if the head end casters are locked along with the central lock and/or
foot end casters.

ELECTRONICS

CONTROL BOX
The picture below shows the various control box connections with actuators, hand pendant
and battery backup.
BED UNDERCARRIAGE
12 FCE1385B (PIC-00391) | Revision: 03/26/20

BATTERY BACKUP

In the event of a power failure, the bed comes standard with a battery backup feature
that enables the bed to be lowered without AC power. Keep the battery backup plugged
into control box to ensure activation. Battery must be charged for a minimum of 12 hours
prior to use.
The battery backup unit will recharge when AC power is reestablished.
CAUTION
Improper handling and/or exposure to heat could result in bodily injury.
The battery back-up is a factory sealed unit. DO NOT ATTEMPT to open/repair unit.
Contains sealed lead battery. Battery must be recycled.
REPLACING BATTERY
Tools Required: Flat Head Screw Driver, T25 Torx Screw Driver
CAUTION
Follow the installation instructions carefully and check that all hardware/fasteners are
properly installed and tight.
1. Disconnect AC power
2. Remove the control box cover by using a flat head screw driver. Insert the screwdriver into
the control box cover slot marked "INSERT TO OPEN". The control box cover should now
slide away from the control box.
3. Unplug the battery backup cable from the control box. Please refer to page 13 for the
battery backup cable location on the control box.
4. Cut and remove all cable (zip) ties holding the battery backup cable.
5. Using a T25 Torx screwdriver remove the all 4 mounting screws. Support the battery when
removing the last screw. Use caution the battery backup weighs approximately 4 pounds.
6. Using a T25 Torx screwdriver install the new battery backup unit with all 4 mounting
screws.
7. Carefully route the battery backup cable from the battery to the control box. Please make
sure the cable does not get pinched or cut during normal bed operation.
8. Secure cable in place with zip-ties
9. Reconnect the battery backup cable into the control box port.
10. Replace the control box cover.
11. Reconnect AC power
Please recycle the original battery backup unit.