OPERATION AND MAINTENANCE MANUAL ALTERRA 1385, LONG TERM CARE BED (Shown with optional side rails, mattress, head/footboards, pillows, and comforter) This product is not made with natural rubber latex. Packaging / Handling Instructions see Page 8. FCE1385B...
FDA has not completed harmonization with REORDER GUIDE ........................Page 19 the measurements in this IEC document. The Medline side rails listed above do meet the dimensional requirements for the Zone 6 area that are defined by IEC60601.
Dimensions The information in this document is subject to change without notice. Check all parts for shipping damage and test before using. DO NOT USE IF DAMAGED. Contact Medline Industries Gaps within the rail (Zone 1) <120 mm (< 4 3/4 “) Inc.
Please have the serial number and model number of the bed available when calling • DO NOT cover the bed’s power cord with rugs, carpet, etc. Medline Customer Service for parts and/or troubleshooting. 6 FCE1385B (PIC-00391) | Revision: 03/26/20 1-800-MEDLINE (633-5463) | medline.com 7...
(this may require a rubber mallet). Use caution as to not damage the ** If at this point, you discover concealed damage, immediately contact Medline Customer surface of the footboard. The surface with the screw inserts visible will be the outside of Service and take photographs to document the damage for the freight claim.
Secure the individual locking casters. To roll the bed, disengage the central lock by pressing the pedal and release the individual caster locks. Activating the folding central lock does not disengage the bed’s functions. 10 FCE1385B (PIC-00391) | Revision: 03/26/20 1-800-MEDLINE (633-5463) | medline.com 11...
8. Secure cable in place with zip-ties 9. Reconnect the battery backup cable into the control box port. 10. Replace the control box cover. 11. Reconnect AC power Please recycle the original battery backup unit. 12 FCE1385B (PIC-00391) | Revision: 03/26/20 1-800-MEDLINE (633-5463) | medline.com 13...
Is the bed plugged into Plug bed Into wall outlet IF USE OF THIS TROUBLE SHOOTING GUIDE DOES NOT CORRECT THE PROBLEM, PLEASE CALL an outlet? MEDLINE TECHNICAL SUPPORT AT 1-800-MEDLINE. Is the power outlet on Turn switch on, check fuse a switch?
6. Plug in the actuator cable. 7. With at least TWO people, carefully return the bed to the upright position. 8. Reconnect AC power. 16 FCE1385B (PIC-00391) | Revision: 03/26/20 1-800-MEDLINE (633-5463) | medline.com 17...
5. Check all bed functions a. Head raises and lowers properly. b. Foot raises and lowers properly. c. Bed raises and lowers properly. d. Check for malfunctions, recalls, or safety notices. 20 FCE1385B (PIC-00391) | Revision: 03/26/20 1-800-MEDLINE (633-5463) | medline.com 21...