2. CyFlow
space Optics
Laser / Output
Detectors
Filters
Video Flow Monitor
Optical Coupling
Excitation Optics
®
3. CyFlow
space Fluidics
Flow Cuvette
Sample Delivery
Sampling Volume
Flow Rates
Fluidics Volume
BioSafety System
®
4. CyFlow
space Electronics and Computer
Electronics
Computer Desktop:
(specifications continiously updated
according to improved performance
of available hardware and software
components)
Rev. 007 Date 2010-01-13
12 months on all parts except filters, mirrors, other quartz or glass parts,
disposables and cuvettes
Red Diode Laser:
Green solid-state Laser:
Blue solid-state Laser:
Violet Diode Laser
Ultra Violet Diode Laser
Laser specifications may vary. Other lasers on request.
1 to 9 (FSC, SSC, FL1, FL2, FL3, FL4, FL5, FL6, FL7)
Standard setup and filters for FSC, SSC, FL1-FL7
Colour CCD camera for video flow monitor
Standard objective mount with high numerical aperture objective, high numerical
aperture immersion gel coupling, e.g. for detection of weak cytokines (option)
Elliptical 15 µm x 70 µm at 488 nm
Other beam geometries upon request
Synthetic quartz flow cuvette (350x 200 µm) for laminar sample transport with
sheath fluid for fluorescence, forward and side scatter light detection
Computer controlled precision syringe pump for contamination-free sample
transport.
Built-in air pressure for sheath fluid. Sheath fluid pressure is adjustable from 0-300
mbar (Computer controlled). Default setting: 200 mBar
Continuous up to 1500 µl
200 µl for precision absolute counting
Other counting volumes upon request
Free sampling volume with syringe counting method
1) Sample volume speed adjustable continously between 0 and 50 µl/s
2) Sheath fluid pressure continuously adjustable
2 x 2-litre reservoirs for sheath fluid and waste
Avoids sample droplets and sample cross contamination (computer controlled)
Parallel realtime signal processing for each of the optical channels with selectable
linear, 3- or 4-decade logarithmic amplification (6-decade log amplifiers), pulse
height, area and width analysis for doublet discrimination, 16 bit analog-to-digital
converters, trigger on any parameter or all parameters
Processor ≥ Pentium 4 3.0 GHz, ≥ 512 MB RAM (or equivalent)
Harddisk ≥ 160 GB
Floppy disk drive 1.44 MB and DVD CD-RW
19" TFT monitor
Keyboard and mouse
Microsoft Windows XP® professional
Microsoft Office® 2007 Basic
25 mW or 40 mW at 638 nm
30 mW to 100 mW at 532nm
20 mW or 200 mW at 488 nm
100 mW at 405 nm
16 mW at 375 nm
®
CyFlow
space Instrument Operating Manual