The Parameter Setup Dialog Box: Pulse Height, Area And Width - Partec CyFlow space Instrument Operating Manual

Table of Contents

Advertisement

Operating Basics

The Parameter Setup Dialog Box: Pulse Height, Area and Width

Opening the
Parameter Setup
Note
Parameter
names
cannot
be
changed
after an acquisition.
Rev. 007 Date 2010-01-13
Fig. 5: The Parameter Setup Dialog Box
Parameter names
Pulse Property
Selections
Low Pass
2 Wavelengths
Signal Delay
Before
starting
an
parameters of interest are selected and set up
according to your analysis.
Click
the
"Parameters..."
Instrument Settings box.
The Parameter Setup Dialog Box appears.
Check if the parameter names are according to your
optical setup. Typical names are shown in the
example
above.
The
displayed on the histogram axis together with a
parameter label. To change parameter names, click
into the fields and enter the names by using the
keyboard.
For each optical channel, you may analyse three
different pulse properties: 1. The pulse height, 2. the
pulse area, and 3. the pulse width. Typically, only
the pulse height is used for analysis. In case of a
DNA analysis, the pulse area of the DNA parameter
may be selected in order to discriminate cell
doublets from single cells, which show up in the
same DNA peak.
The electrical low pass filter smoothes the signals
from the detectors by averaging the signal over a
given time. This feature is only required for lamp
light sources (not realized for the CyFlow
Light beams from different lasers are directed to
different positions (spots) of the flow cuvette (please
refer to page 6). Therefore, all laser derived signals
from a laser at a spot other than spot no. 1 need to
be analyzed with a time delay of 50 µsec (2-
Wavelengths Signal Delay activated, see Fig. 5)
®
CyFlow
space Instrument Operating Manual
acquisition,
assure
button
in
parameter
names
®
space).
the
the
are
13/26