Cardinal Health ALLY TO GO NPWT Patient User Manual page 15

Negative pressure wound therapy
Table of Contents

Advertisement

2. Canister Full/Blockage - With Clamp on Tubing
(if no clamp, see next instructions)
WHAT YOU
SEE/HEAR
WHAT TO DO
First, visually assess the canister to see if it is full. If the canister is full, change the canister (2.3 Changing the Canister).
If the canister is not full, proceed with the steps below.
Make sure
1
clamp is
OPEN.
Wait
4
If alert
30 seconds
stays on, go
to step 5.
7
Replace with new canister,
reattach tubing and open clamp.
If you see:
9
SOLID LIGHT
Canister Full/Blockage is fixed.
NOTE: A false Canister Full/Blockage alert can occur if drainage moves into the canister while the ALLY TO GO™ is
lying on its front (the button side) or if it is turned upside down. If this occurs, you must change the canister to
resume negative pressure wound therapy. Always keep the ALLY TO GO™ upright.
+
BLINK
TWO BEEPS
Wait
If alert
2
30 seconds
continues,
go to step 3.
5
Turn OFF.
If you see/hear:
+
BLINK
Look for kinks in
3
tubing. Unkink
if found.
6
Clamp tubing and remove
canister
(2.3 Changing the
.
Canister)
8
Turn ON.
+
+
BLINK
TWO BEEPS
Call Cardinal Health at 1.866.484.6798
Wait
1 minute
BLINK
15