This document describes the preferred calibration procedure of the WatchGas UNI product range. If you require
any assistance, feel free to contact us at info@watchgas.nl
Equipment
To calibrate the WatchGas QGM the follow equipment is needed:
WatchGas QGM
WatchGas QGM Calibration adapter
0.5 L/min regulator
Last-O-More hose
WatchGas IR link (optional)
Calibration gas
The use of the WatchGas IR link and IR link software is not required, but it's certainly recommended. Span
values can only be changed using the IR link software.
The WatchGas QGM can only measure O
gases. Note that these gases are available in a multi-gas mixture. Calibrating using a multi-gas mixture is
most cost and time efficient.
Adequate safety measures should be taken when working with these type of gases. Avoid personal
exposure at all times by working under a fume hood and/or wearing personal protective equipment.
List with commonly used calibration gases
Description
18%O
/ 2.2%CH
/ 50CO / 10H
2
4
18%O
/ 2.2%CH
/ 50CO / 10H
2
4
18%O
/ 2.5%CH
/ 2%CO
/ 50CO / 15H
2
4
2
100ppm I-C
H
in Air
4
8
100ppm I-C
H
in Air
4
8
18%O
/ 2.2%CH
/ 50CO / 10H
2
4
18%O
/ 2.2%CH
/ 0.5%CO
2
4
2
18%O
/ 2.5%CH
/ 100CO / 25H
2
4
100% N
(UHP)
2
18%O
/ 2.2%CH
/ 0.5%CO
2
4
2
100ppm I-C
H
in Air
4
8
50ppm NH
in N
3
2
20.9% / 1.1%CH
/ 200ppm CO in N
4
15%O
/ 1.45%CH
/2.5%CO
2
4
2
18%O
/ 2.2%CH
/ 100CO / 25H
2
4
15%O
/ 1.45%CH
/ 60CO / 20H
2
4
18%O
/ 2.2%CH
/ 100CO / 25H
2
4
18%O
/ 2.2CH
/ 50CO in N
2
4
2
10ppm H
S in N
2
2
25ppm H
S in N
2
2
, LEL, CO and H
2
S in N
2
2
S in N
2
2
S in N
2
2
S in N
2
2
/ 100CO / 25H
S in N
2
2
S in N
2
2
/ 100CO / 25H
S in N
2
2
2
/ 60CO / 20H
S in N
2
2
S in N
2
2
S in N
2
2
S in N
2
2
WatchGas QGM Calibration Manual
S. Please refer to our well-priced list of calibration
2
Size
116L
34L
58L
34L
116L
58L
58L
34L
116L
34L
58L
58L
116L
58L
58L
34L
34L
116L
116L
58L
Tech/App Note 6 part 2
Article Number
CAL-C017064
CAL-C009147
CAL-C015013
CAL-C008920
CAL-C006392
CAL-C011916
CAL-C012017
CAL-A1069841
CAL-A0961956
CAL-C009266
CAL-C015004
CAL-C011793
CAL-C019604
CAL-C012032
CAL-C011921
CAL-C009154
CAL-A0915728
CAL-A0975762
CAL-C017065
CAL-C011126

Advertisement

loading

Summary of Contents for WatchGas QGM

  • Page 1 WatchGas IR link (optional) • Calibration gas The use of the WatchGas IR link and IR link software is not required, but it’s certainly recommended. Span values can only be changed using the IR link software. The WatchGas QGM can only measure O , LEL, CO and H S.
  • Page 2 Manual calibration using the IR link To calibrate the WatchGas QGM using the IR link, use the following steps: 1. Turn on the device and wait for all sensor to warm-up. 2. Place the device onto the IR link and start the IR link software 3.
  • Page 3 Automatic calibration using the WatchGas QGM Docking station To calibrate the WatchGas QGM with Docking Station, use the following steps: 1. Turn on the device and wait for all sensor to warm-up. 2. Turn on the docking station. Ensure the docking station is configured correctly and there’s sufficient gas remaining on the internal gas cylinder.
  • Page 4 Manual calibration To calibrate the WatchGas QGM by hand, use the following steps: 1. Turn on the device and wait for all sensor to warm-up. 2. Press the single button 8 times to reach the calibration menu, indicated by the gas cylinder icon in the upper right corner.