Manual EICKEMEYER ® NarkoVet Table of Contents Description ...................................2 Overview of the Device ..............................2 Startup Procedure ................................3 Operation of the Pendulum System (Bain Circuit) ....................5 Preparing the Device before Anesthaesia ........................6 Anaesthetic Procedure ..............................6 Cleaning and Maintenance of Anesthaesia Equipment ..................7...
Manual 1 Description The anesthaesia machine NarkoVet is a semi-closed loop system with partial rebreathing. The device is suitable for anesthaesia of small animals up to a body weight of 80 kg. For animals up to 15 kg it is recommended to use the included pendulum system (Bain Circuit).
Thus, the filling time of the breathing bag is extended when operating the flush button (oxygen bypass (2)). The flow meter, however, always displays the correct oxygen flow. NarkoVet with 10 l oxygen cylinder and pressure reducer (not included in the delivery) Fill the evaporator with Isoflurane.
Page 6
EICKEMEYER ® NarkoVet Manual Filling the soda lime Turn the lock lever of the absorbent container (11) counterclockwise. Remove the soda lime. Fill the container with soda lime. The max. filling line must not be exceeded. Insert the container and lock it by turning the lock lever clockwise.
EICKEMEYER ® NarkoVet Manual 4 Operation of the Pendulum System (Bain Circuit) The Bain Circuit supplements the basic unit perfectly for anesthaesia of smaller patients. It contains no breathing valves. The anesthetic gas is transported directly from the evaporator via the fresh gas pipe to the patient. This ensures quicker wash-in and wash-out even with a smaller minute volume.
EICKEMEYER ® NarkoVet Manual 5 Preparing the Device before Anesthaesia Check the oxygen supply (pressure and access to sufficient oxygen). Connect patient tube and breathing bag to the anesthesia machine and close the pressure relief valve. Keep your thumb on the patient's end of the tube and inflate the breathing bag via the Oxygen bypass until the breathing bag is bulging.
Once a year, the evaporator should be subjected to a technical inspection to ensure proper functioning. In this case, please contact the EICKEMEYER® Customer Service. Precautions for handling Isoflurane: Isoflurane is a highly volatile anaesthetic. Therefore, when used in closed spaces ensure that a suction tube is properly connected to the pressure relief valve or the Bain Circuit.