Please read this manual thoroughly before use and use the product in a correct way. Please carefully keep the User Manual for future use. EICKEMEYER® is responsible for the reliability and performance of the equipment only all following conditions are met: •...
If there’s any abnormal phenomenon for the puncturing part, please timely take suitable measures, such as puncturing again. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
fl ow rate, the longer the time of detecting occlusion, and it in turn may possibly cause a long time infusion stop during this period. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
The limited humidity range of transportation Protective earth (ground) package is 10 ~ 95 % The environmental pressure of transportation package is limited to 50 ~ 106 kPa Table 1.4-1 T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
This equipment provides several infusion modes, such as ml/h mode, body weight mode, drip mode, sequence mode. Additionally, it also has functions such as history records, drug library, Anti-bolus, and alarm and so on. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Minimum step is 0.01 ml/h < 100 ml/h step is 0.01 ml/h < 1,000 ml/h step is 0.1 ml/h ≥ 1,000 ml/h step is 1 ml/h T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Page 13
100 – 240 V AC, 50 / 60 Hz, 0.25 – 0.1 A Input power 50 VA DC power supply DC 10 – 16 V, 1.5 – 0.94 A T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
⑧ Anti-free fl ow clamp ⑨ Display screen ⑩ Handle Door switch Note! It is recommended that the waterproof fi lm be replaced once every two years. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
⑧ Alarm indicator (red / yellow) While pump alarms, indicator light glitters, with different frequency and color, more information please refer to Chapter 8.1. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Since the remained battery may change, it may possibly Battery status indication icon show the following states: Table 3.3.1-1 T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Click it to clear input Backspace button Click it to backspace delete Toggle key Click to switch into English or case input setting Table 3.3.2-1 T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
③ ① Housing ② Slider Push the slider to left direction to adjust the spacing, loosen the slider to automatically return. ③ Cable T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Ensure the stability of the animal cage that is attached to the infusion pump. Be careful when moving the animal cage and this equipment to prevent the equipment from slipping or colliding with nearby objects. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
The fl uid / drug volume in the murphy’s dropper must be less than 1 / 3 of its volume. • The drop sensor shall be vertical. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
It may possibly drop the infusion accuracy or abnormal work of the equipment if the working environment temperature exceeds the designated range. • The viscosity and specifi c gravity of infusion fl uid will infl uence the infusion accuracy. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Select “Last treatment” to enter the last usage mode parameter setting interface. Warning! If the self-test item does not pass, please contact EICKEMEYER® and you are not allowed to continue using the equipment. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
4. Close the pump door, then pop up the infusion tube selection interface, indicating that the infusion tube is installed correctly. Otherwise you need to reinstall. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Warning! EICKEMEYER® is the brand of infusion set built in the device. If using a non-built-in infusion set, please confi rm the relevant infusion performance (accuracy, air bubble, pressure) on the infusion pump before confi rming the use, otherwise the infusion will not be guaranteed.
During the infusion process, click the fl ow rate value on the running interface to reset the fl ow rate. After confi rmation, click to continue the infusion. Note! Sequence mode does not support changing the fl ow rate during infusion. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
3. Cancel: click『Cancel』, return to the interface before OFF setting. 4. If no operation, the device will enter standby interface automatically. Note! The equipment has standby function only under the non-working state. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
The higher the chosen click level, the higher the occlusion level, it is suggested to select suitable occlusion pressure according to actual requirement. DPS is turned on by default, and the line pressure is graphically and dynamically visible during infusing status. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Click『Weight unit』to enter into the body weight unit setting interface, click preset body weight unit option, then the corresponding icon of this option changes into Note! The current software version only support unit kg. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
The infusion apparatus of different brand may possible cause fl ow rate deviation, when use, please confi rm if the displayed information in the interface is accordant with the actual working infusion apparatus. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
If the screen or keypad is locked, no operation can be conducted. After turn on 『Screen lock』function during infusing, press【Power】key to lock or unlock the device manually. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
2. The system can save 20 medical records at most, when it is full, the new records will cover the old records by turn. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
10 years. When the power failure time is ≤ 30 s, the alarm setting before power failure will be automatically recovered. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
When there’s alarm, please check the conditions of the animal, remove the reason of alarm and then continue working. Please refer to Appendix D for the alarm solution. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Please eliminate the fault alarm according to the prompt. If it can’t be eliminated, please stop the equipment, contact EICKEMEYER® to repair and test the equipment, do not put it into operation before the equipment has passed the inspection, otherwise, it may possibly cause unpredictable harm if it works with fault.
6 months. Some items shall be maintained by the user, and some items shall be maintained by the dealer of the equipment. • Please timely contact EICKEMEYER® if the equipment is found defective. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Battery performance. 9.2.3 Maintenance Plan The following check / maintenance items must be performed by the professional technician recognized by EICKEMEYER®. If the following maintenances are necessary, please contact EICKEMEYER®. Please clean and disinfect the equipment before testing or maintaining.
Page 39
Note! • When the『Volume』is less than 10 ml, the infusion rate is ≤ 1,500 ml/h. • When the『Volume』is less than 7.5 ml, the infusion rate is ≤ 1,200 ml/h. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
9.4.1 Normal Repair Process Please contact EICKEMEYER® or authorized service personnel to repair if there’s any fault, do not disassemble and repair the equipment. After repair, please perform overall test for the equipment. EICKEMEYER® may provide the circuit diagram and components list to the authorized repair technician if needed.
10.2 Replaced the Battery It is recommended to replace the battery every 2 years, it is suggested to replace the battery by the dealer or EICKEMEYER®. Warning! Untrained personnel are forbidden to replace the battery, otherwise it may cause the battery to burn, explode, leak and cause personal injury.
3. The fault and damages caused by fi re, salt injury, toxic gas, earthquake, windstorm, fl ood, abnormal voltage and other natural disasters. For the damages or faults mentioned above, EICKEMEYER® provides repair services but charge at repair cost. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
The above data is the typical value under the test conditions, please see the test data of the product for the actual data, the data may be different if the test conditions are different T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Restart device to check whether alarm is System error High system error alarm will give with code eliminated, if still exists, contact number. maintenance personnel. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Page 45
Note! When alarm rings, click the『Mute』icon on the screen to temporarily stop sound alarm for 2 min. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
The use of accessories, transducers and cables other than those specifi ed, with the exception of transducers and cables sold by EICKEMEYER® of the Infusion pump as replacement parts for internal components, may result in increased emissions or decreased immunity of the Infusion pump.
Page 47
fi eld IEC 61000-4-8 location in a typical commercial or hospital environment. NOTE is the a.c. mains voltage prior to application of the test level. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Page 48
Infusion pump. Over the frequency range 150 kHz to 80 MHz, fi eld strengths should be less than 10 V/m. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Page 49
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and refl ection from structures, objects and people. T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Brightness 50 % Anti-bolus Bubble size 100 µl Cumulative Bubble 300 µl / 15 min Screen lock 1 min Occlusion pressure 600 mmHg Finish pre-alarm 2 min Reminder alarm 2 min T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Page 51
USER MANUAL INFUSOVET CONNECT INFUSION PUMP NOTES T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Page 52
GERMANY SWITZERLAND POLAND EICKEMEYER KG EICKEMEYER AG EICKEMEYER Sp. z o.o. Eltastraße 8 Sandgrube 29 Al. Jana Pawła II 27 78532 Tuttlingen 9050 Appenzell 00-867 Warszawa T +49 7461 96 580 0 T +41 71 788 23 13 T +48 22 185 55 76...