Eickemeyer ORTHOVET PLUS User Manual

Eickemeyer ORTHOVET PLUS User Manual

Battery drill system

Advertisement

Quick Links

ORTHOVET PLUS
BATTERY DRILL SYSTEM
USER MANUAL
Item No. 195070
TELEPHONE +49 7461 96 580 0
www.eickemeyer.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ORTHOVET PLUS and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Eickemeyer ORTHOVET PLUS

  • Page 1 ORTHOVET PLUS BATTERY DRILL SYSTEM USER MANUAL Item No. 195070 TELEPHONE +49 7461 96 580 0 www.eickemeyer.com...
  • Page 2 USER MANUAL ORTHOVET PLUS BATTERY DRILL SYSTEM T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
  • Page 3: Table Of Contents

    This description alone does not provide suffi cient background for the direct use of the product. Instruction in handling this product is highly recommended. T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
  • Page 4: Introduction

    The OrthoVet Plus is only to be used for patient treatment after careful consultation of the instructions for use. It is recommended that an alternative system is available to use during application, as technical problems can never be completely ruled out.
  • Page 5 • The batteries may never be sterilized, washed or dipped. Sterilization would destroy the battery with possible secondary damage. • Should the machine drop on the fl oor and have visible defects, do not use it anymore and send it to the EICKEMEYER® service center.
  • Page 6: Orthovet Plus

    3 Trigger for speed regulation 4 Cover lid Battery closed when the battery is inserted 5 Release ring for attachment 6 Release slide for battery housing T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
  • Page 7: Use

    The same person presses it completely into the battery casing to ensure a correct seat (Fig. 9). This person may not contact the outside of the battery casing. Figure 8 Figure 9 T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
  • Page 8 Storing When not in use, always store the battery in the OrthoVet Plus charging station and switch it on. This prevents the battery from being discharged and keeps it ready for use with full charge. Never load batteries in another charging station. This would damage the batteries.
  • Page 9: Attachments / Chucks

    Please read the guidelines for the transport of nickel-metal hydride (NiMH) for the return transport to the EICKEMEYER® Service Center. 3. ATTACHMENTS / CHUCKS General Information The OrthoVet Plus System offers a broad range of attachments. All attachments has color-coded rings, so that they can easily be identifi ed. Mounting the Attachments 195079...
  • Page 10 • To prevent injury, the power tool must be switched “OFF” with the safety system during each manipulation. • Only use original attachments and tools from EICKEMEYER®. Damage that arises from using attachments and tools made by other manufacturers is not covered by the warranty.
  • Page 11: Jacobs Chuck (Item No. 195078)

    • Use the Jacobs Chuck with Reverse motion only with tools that are approved for such use. Otherwise, the tool may break with subsequent damage. T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
  • Page 12: Quick Chuck For Kirschner-Wires (Item No. 195079)

    6. To adjust the grip on the wire, reduce the clamping force and move the tool to the desired length. Reclamp the wire by pulling on the tension lever. Figure 16 Oscillating Saw-Chuck (Item No. 195081) For saw blades with Aesculap-Connection. Figure 17 T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
  • Page 13 Figure 20 Precaution To prevent injury, always grip the saw attachment with the inserted saw blade from the direction of the machine. T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
  • Page 14 Very precise cuts can be made when the saw blade is guided steadily. Imprecise cuts arise due to used blades, excess pressure or jamming the saw blade. T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
  • Page 15: Ao-Quick Chuck (Item No. 195080)

    Handpiece, funnel and chucks / attachments may be processed using: a) Manual cleaning and / or b) Automated cleaning with manual pre-cleaning Note: Clean all movable parts in opened position. T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
  • Page 16: Manual Cleaning Instruction

    Final rinse with de-ionized or purifi ed water for a minimum of 2 minutes. 8. Dry Dry device using a soft lint-free cloth or clean compressed air. T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
  • Page 17: Mechanical / Automated Cleaning Instruction With Manual Pre-Cleaning

    6. Visually Inspect Device Inspect the cannulations, coupling sleeves, etc. for visible soil. Repeat steps 1 – 6 until no visible soil remains. T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
  • Page 18 / disinfection is an additional stress for power equipment, especially for seals and bearings. Therefore, systems must be properly lubricated and regularly sent to be serviced (at least once per year). T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
  • Page 19: Lubrication

    • Lubricate the attachment release buttons and then press the buttons several times. • Lubricate the mode selector switch and then move it several times. • Lubricate the attachment coupling. Figure 28 T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
  • Page 20 Hold the Quick Coupling up and add one drop of oil into the attachment hole and on the holder of the lever (Fig. 31). Move the tension lever several times. Figure 31 T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
  • Page 21 • To ensure a long service life and reduce repairs the power tool and all attachments must be lubricated after each use. • Only lubricate the power tool and attachments when clean. T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
  • Page 22: Function Control

    Packaging Put cleaned, dry products into the proper places in the EICKEMEYER® tray (item no. 195086) and put the tray into the container (item no. 185555). Care should be taken in order to protect implants as well as pointed or sharp instruments from contact with other objects that may damage the surface or the Sterile Barrier System.
  • Page 23: Repairs And Technical Service

    Repairs and Technical Service • The tool should be sent to the EICKEMEYER®s offi ce for repair if it is faulty or malfunctions. • Contaminated products have to run through the complete reprocessing procedure before being sent to the EICKEMEYER®...
  • Page 24: Troubleshooting

    If the recommended solutions do not work, send the power tool back to EICKEMEYER® service center. For further technical questions or information on our services, please contact your EICKEMEYER® representative. T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
  • Page 25: System Specifications

    Depending on the cutting tool used and on the load applied, the heat generation of the handpiece, attachment and / or cutting tool can vary. Always control the temperature of the system to prevent overheating and possibly harming the patient or user. T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
  • Page 26: Ordering Information

    Saw blade Aesculap-connection, Dimensions (in mm): L 45 x W 10 x H 0.5 196160 Saw blade Aesculap-connection, Dimensions (in mm): L 36 x W 12 x H 0.5 T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
  • Page 27 USER MANUAL ORTHOVET PLUS BATTERY DRILL SYSTEM NOTES T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
  • Page 28 GERMANY POLAND EICKEMEYER KG EICKEMEYER Sp. z o.o. Eltastraße 8 Al. Jana Pawła II 27 78532 Tuttlingen 00-867 Warszawa T +49 7461 96 580 0 T +48 22 185 55 76 F +49 7461 96 580 90 F +48 22 185 59 40 info@eickemeyer.de...

This manual is also suitable for:

195070