This description alone does not provide suffi cient background for the direct use of the product. Instruction in handling this product is highly recommended. T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
The OrthoVet Plus is only to be used for patient treatment after careful consultation of the instructions for use. It is recommended that an alternative system is available to use during application, as technical problems can never be completely ruled out.
Page 5
• The batteries may never be sterilized, washed or dipped. Sterilization would destroy the battery with possible secondary damage. • Should the machine drop on the fl oor and have visible defects, do not use it anymore and send it to the EICKEMEYER® service center.
3 Trigger for speed regulation 4 Cover lid Battery closed when the battery is inserted 5 Release ring for attachment 6 Release slide for battery housing T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
The same person presses it completely into the battery casing to ensure a correct seat (Fig. 9). This person may not contact the outside of the battery casing. Figure 8 Figure 9 T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Page 8
Storing When not in use, always store the battery in the OrthoVet Plus charging station and switch it on. This prevents the battery from being discharged and keeps it ready for use with full charge. Never load batteries in another charging station. This would damage the batteries.
Please read the guidelines for the transport of nickel-metal hydride (NiMH) for the return transport to the EICKEMEYER® Service Center. 3. ATTACHMENTS / CHUCKS General Information The OrthoVet Plus System offers a broad range of attachments. All attachments has color-coded rings, so that they can easily be identifi ed. Mounting the Attachments 195079...
Page 10
• To prevent injury, the power tool must be switched “OFF” with the safety system during each manipulation. • Only use original attachments and tools from EICKEMEYER®. Damage that arises from using attachments and tools made by other manufacturers is not covered by the warranty.
• Use the Jacobs Chuck with Reverse motion only with tools that are approved for such use. Otherwise, the tool may break with subsequent damage. T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
6. To adjust the grip on the wire, reduce the clamping force and move the tool to the desired length. Reclamp the wire by pulling on the tension lever. Figure 16 Oscillating Saw-Chuck (Item No. 195081) For saw blades with Aesculap-Connection. Figure 17 T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Page 13
Figure 20 Precaution To prevent injury, always grip the saw attachment with the inserted saw blade from the direction of the machine. T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Page 14
Very precise cuts can be made when the saw blade is guided steadily. Imprecise cuts arise due to used blades, excess pressure or jamming the saw blade. T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Handpiece, funnel and chucks / attachments may be processed using: a) Manual cleaning and / or b) Automated cleaning with manual pre-cleaning Note: Clean all movable parts in opened position. T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Final rinse with de-ionized or purifi ed water for a minimum of 2 minutes. 8. Dry Dry device using a soft lint-free cloth or clean compressed air. T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
6. Visually Inspect Device Inspect the cannulations, coupling sleeves, etc. for visible soil. Repeat steps 1 – 6 until no visible soil remains. T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Page 18
/ disinfection is an additional stress for power equipment, especially for seals and bearings. Therefore, systems must be properly lubricated and regularly sent to be serviced (at least once per year). T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
• Lubricate the attachment release buttons and then press the buttons several times. • Lubricate the mode selector switch and then move it several times. • Lubricate the attachment coupling. Figure 28 T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Page 20
Hold the Quick Coupling up and add one drop of oil into the attachment hole and on the holder of the lever (Fig. 31). Move the tension lever several times. Figure 31 T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Page 21
• To ensure a long service life and reduce repairs the power tool and all attachments must be lubricated after each use. • Only lubricate the power tool and attachments when clean. T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Packaging Put cleaned, dry products into the proper places in the EICKEMEYER® tray (item no. 195086) and put the tray into the container (item no. 185555). Care should be taken in order to protect implants as well as pointed or sharp instruments from contact with other objects that may damage the surface or the Sterile Barrier System.
Repairs and Technical Service • The tool should be sent to the EICKEMEYER®s offi ce for repair if it is faulty or malfunctions. • Contaminated products have to run through the complete reprocessing procedure before being sent to the EICKEMEYER®...
If the recommended solutions do not work, send the power tool back to EICKEMEYER® service center. For further technical questions or information on our services, please contact your EICKEMEYER® representative. T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Depending on the cutting tool used and on the load applied, the heat generation of the handpiece, attachment and / or cutting tool can vary. Always control the temperature of the system to prevent overheating and possibly harming the patient or user. T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Saw blade Aesculap-connection, Dimensions (in mm): L 45 x W 10 x H 0.5 196160 Saw blade Aesculap-connection, Dimensions (in mm): L 36 x W 12 x H 0.5 T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Page 27
USER MANUAL ORTHOVET PLUS BATTERY DRILL SYSTEM NOTES T +49 7461 96 580 0 | F + 49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com...
Page 28
GERMANY POLAND EICKEMEYER KG EICKEMEYER Sp. z o.o. Eltastraße 8 Al. Jana Pawła II 27 78532 Tuttlingen 00-867 Warszawa T +49 7461 96 580 0 T +48 22 185 55 76 F +49 7461 96 580 90 F +48 22 185 59 40 info@eickemeyer.de...