Page 1
CS 7600 Safety, Regulatory and Technical Specifications...
Page 2
Make sure you have the most current version. No part of this Guide may be reproduced without the express permission of Carestream Dental, LLC. U.S. Federal law restricts this device to sale by or on the order of a dentist or physician.
Safety Information Indications for Use The CS 7600 is intended for digital dental radiography using an imaging plate (storage phosphor screen) for radiographic diagnostic intraoral images. Conventions in this Guide The following special messages emphasize information or indicate potential risk to personnel or equipment.
Warning and Safety Instructions The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally required) this equipment might not offer adequate protection to radio frequency communication services.
Page 7
DO NOT use this device in the presence of a flammable anesthetics mixture with air or with oxygen or with nitrous oxide. DO NOT remove or open system covers or plugs. Internal circuits use high voltage capable of causing serious injury. Fuses blown within 36 hours of being replaced by a qualified technician may indicate malfunctioning electrical circuits within the system.
Page 8
Disconnecting items from the IT-NETWORK. Update of equipment connected to the IT-NETWORK. Upgrade of equipment connected to the IT-NET. Imaging Plates: To prevent damage to the imaging plates and the possibility of image artifacts, avoid contact between the imaging plates and the following materials/solutions/solvents: Isopropyl alcohol, hydrogen peroxide and other peroxides, citrus-based cleaners, hand lotions and water-less hand sanitizers, as well as surfactants and lubricants.
Page 9
See your computer installation guide for details of the data processing system and screen. Leave a sufficient amount of clear space around the CPU to ensure that it is properly ventilated. To obtain maximum image quality and visual comfort, position the screen to avoid direct light reflections from internal or external lighting.
General Notice WARNINGS If the product does not operate properly or fails to respond to the controls as described in the product accompanying documentation: Follow the safety precautions as specified in this guide. Stop using the unit and prevent any changes to it. ...
Page 11
WARNING Do not operate the equipment in the presence of explosive liquids, vapors, or gases. Do not plug in or turn on the system if hazardous substances are detected in the environment. If these substances are detected after the system has been turned on, do not attempt to turn off the unit or unplug it.
Warning: General warning sign Using accessories other than those specified in this document with the exception of those sold by Carestream Dental LLC may result in a lower level of security for the entire system. General mandatory action sign. Refer to instruction manual / booklet.
IEC Symbols Used The system may have labels with one or more of the following symbols Label Description Caution—consult accompanying documents Power On/Off Caution—Electrical shock hazard Safety Labels Label Description Location Caution—Electrical PM board shock hazard Laser-emitting product Optical head Class 3B laser product Optical head and on the scanner inside scanner...
The following figure illustrates the label locations of the CS 7600 components. Scanner FCC ID: 2ASFG-CR-SCANNER IC: 24914-CRSCANNER Manufacturer: Carestream Dental LLC This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:...
Regulatory Information WARNING: Any changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate this equipment. General Regulatory Information The CS 7600 is an active device specifically intended for scanning of X-ray diagnostic images.
Page 16
Compliance with European and International Standards Compliance with European and International Standards EN 60601-1 / IEC Medical Electrical Equipment, Part 1: General 60601-1 Requirements for Basic Safety and Essential Performance Medical Electrical Equipment, Part 1-2: General EN 60601-1-2 / IEC requirements for basic safety and essential performance - 60601-1-2 Collateral Standard: Electromagnetic Disturbances –...
Page 17
Classification in Accordance with EN/IEC 60601-1 Type of protection against Class 1 equipment electric shock Degree of protection against No Applied Parts electric shock Protection against harmful Ordinary equipment, IPX0 ingress of water Operation mode Continuous operation Electromagnetic Compatibility Precautions Conformity with EN/IEC 60601-1-2 Electromagnetic Compatibility Precautions ...
Page 18
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions (IEC 60601-1-2) The CS 7600 is intended for use in the electromagnetic environment specified below. The customer or the user of the CS 7600 should assure that it is used in such an environment. Emissions Test Compliance Electromagnetic Environment - Guidance...
Page 19
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity (Continued) ± ± 1 kV line(s) to 1 kV line(s) to Mains power quality should be Surge line(s) line(s) that of a typical commercial or IEC 61000-4-5 ± ± 2 kV line(s) to 2 kV line(s) to hospital environment.
Page 20
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity (IEC 60601-1-2) The CS 7600 is intended for use in the electromagnetic environment specified below. The customer or the user of the CS 7600 should assure that it is used in such an environment. Immunity IEC 60601 Compliance...
Page 21
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity For the immunity to proximity fields from RF wireless communications equipment, the CS 7600 is compliant with the test levels specified below according to IEC60601-1-2 standard. The customer or the user of the CS 7600 should assure that it is used in such an environment.
Note: The communication without disruption has been determined to be essential performance with regard to electromagnetic compatibility. Compliance with FCC Rules This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: This device may not cause harmful interference.
Technical Specifications Model CS 7600 Factory Address Rayco (Shanghai) Medical Products Company Limited Building 7, No. 1510 Chuanqiao Road, China (Shanghai) Pilot Free Trade Zone 201206 Shanghai China PEOPLE'S REPUBLIC OF CHINA Manufacturer’s Address CS 7600_Safety, Regulatory & Technical Specifications (8J4070)_Ed07...
CS 7600 Technical Specifications System Specifications Components CS 7600 Dimensions 267 mm (H), 237 mm (W), 260 mm (D) (without bracket) Weight Approximately 6 kg SL Power Model #: MW174KB2403F01 100 – 240 V (ac), 50/60 Hz, 1.5 A Power supply* Mean Well Model #: GSM60A24 100-240V(ac), 50/60 Hz, 1.4-0.7A Laser wavelength 635 –...
RF Information RF Information Scan & Go standard ISO15693 Frequency 13.56 MHz CS 7600 Environmental Requirements Ambient Operating Conditions Ambient Operating Conditions Temperature 5 – 35 °C Relative humidity 30 – 85 % Non-condensing Atmospheric pressure 700 – 1060 hPa Storage and Transport Conditions Ambient Storage Conditions (in packaging) Temperature...
Minimum Computer System Requirements Item Viewing and Acquisition Comments For best performance it is 2 GHz Intel® Dual Core™ recommended to use an Intel® Dual Core™ AMD Athlon or higher processor. RAM has a major impact on 1 GB ( 2 GB recommended ) system performance.
Page 29
Authorized Representative in the European Community EC REP Trophy 4, Rue F. Pelloutier Croissy-Beaubourg 77435 Marne-la-Vallée, Cedex 2 France Importer for European Union Carestream Dental Germany GmbH Hedelfinger Str. 60 70327 Stuttgart Germany CS 7600_Safety, Regulatory & Technical Specifications (8J4070)_Ed07...
Page 30
Authorized Representative in Brazil CARESTREAM DENTAL BRASIL EIRELI Rua Romualdo Davoli, 65 1º Andar, Sala 01 - São José dos Campos São Paulo - Brazil CEP (Zip code): 12238-577 Chapter 4 Contact Information...