7.4.1 Inspecting the sensor

If the cause of the fault is suspected to be in the measuring chamber (sensor), the following checks can be made
with an ohmmeter.
Tools / Material Required
• Allen wrench AF 2
• Pliers for retaining ring
• Ohmmeter
3
2
1
 Unfasten the hexagon socket set screw (4) and remove the complete measuring chamber (3) from the
electronics unit (5).
 Remove the retaining ring (1) as well as the ionization chamber (2) from the measuring chamber (3).
 Check the ionization chamber and the measuring chamber for contamination:
• Ionization chamber is contaminated only: Replace ionization chamber.
• Measuring chamber is severely contaminated: Replace complete measuring chamber.
 Using an ohmmeter, make following measurements on the contact pins.
Measurement
between pins
5 + measuring
∞ <<∞
chamber
3
4
2
1
5
Replace measuring chamber if defective.
 It is recommend to perform a leak test (leak rate <1×10
InstruTech
5
4
Possible cause
Contamination,
short circuit
cold cathode
sensor
Measuring chamber
-9
mbar l/s).
CCM502 Hornet
Page 19