Instrutech Hornet CCM502 User Manual page 24

Cold cathode ionization vacuum gauge
Table of Contents

Advertisement

Instruction Manual
 Slide a new ionization chamber (2) into the
measuring chamber (3) until the mechanical stop is
reached and mount the retaining ring (1).
 It is recommend to perform a leak test (leak rate <1×10
 Carefully slide the measuring chamber (3) (clean or new) into the electronics unit (5) until the mechanical
stop is reached.
Pins aligned
straight.
 Fasten the measuring chamber (3) by means of the hexagon socket set screw (4).
InstruTech
-9
mbar l/s).
CCM502 Hornet
Page 21