7.4.3 Replacing the measuring chamber

 Set the calibration value of the spare sensor with the <CAL> switch on the electronics unit (5).
0 2
E
C
4
A
6
8
 Carefully slide the measuring chamber (3) into the electronics unit (5) until the mechanical stop is reached.
Pins aligned straight.
3
 Fasten the measuring chamber (3) by means of the hexagon socket set screw (4).
 It is recommend to perform a leak test (leak rate <1×10
leak detector.
WARNING! Helium may cause electric arcing with detrimental effects on the electronics of the product.
Before performing any leak tests disconnect power and remove the electronics unit.