Elis FLONEX FXx11x Series Control Instructions

Elis FLONEX FXx11x Series Control Instructions

Electromagnetic flowmeters

Advertisement

Quick Links

Project design, installation and service manual
Electromagnetic flowmeters of the type series
Page 1 o
33
f
FLONEX FXx11x and FLONET FH30xx
ELIS PLZEŇ a. s.
Meter control instructions
Electromagnetic flowmeters of the type series
FLONEX FXx11x
and
FLONET FH30xx
Meter control instructions
ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722
Es 90 686 K

Advertisement

Table of Contents
loading

Summary of Contents for Elis FLONEX FXx11x Series

  • Page 1 ELIS PLZEŇ a. s. Meter control instructions Electromagnetic flowmeters of the type series FLONEX FXx11x FLONET FH30xx Meter control instructions ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 2: Table Of Contents

    Frequency for Q− ........................18 5.1.4.4.2. 5.1.4.4.3. Frequency for |Q| ......................... 18 5.1.4.4.4. Impulses for Q+ ........................... 19 ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 3 5.2.1. Flowmeter connection to a communication line .................. 29 5.2.2. Program FLOSET 4.0 installation and operational start ..............30 6. DOCUMENTATION AND STANDARDS ..................32 ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 4: Basic Information

    The meter sensor shall be fully flooded with the measured liquid. More detailed description of the required initial inspection procedure is included in the flowmeter manual. ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 5: Meter Energising

    Button positions with respect to the display unit ▲ ► ⃝ ⃝ ■■■■■■■■■■■■■■■■■■■ 1,025 m3/h 45,34 - 25,1 ⃝ ⃝ ENTER ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 6: Flowmeter Control

    This applies to situations such as ADC failure, open or short-circuited excitation, not fully flooded piping, strong electromagnetic interference and others. ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 7: Error And Diagnostic Messages

    Should no control button (▲, ►, ESC, ENTER) be actuated during the recent three minutes, or should the specified backlighting time elapse before that, the display will return automatically to the Basic screen format. ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 8: Main Menu

    (plant) control system, it suffices to set the parameters shown in bold print in the above list of the Main menu items. ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 9 ELIS PLZEŇ a. s. Meter control instructions Control menu structure of flowmeters FLONEX FXx11x and FLONET FH30xx ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 10: Language

    To move upward in the menu, use button ▲ 100% The selected item will be highlighted and can be open using ENTER. ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 11: Display Data

    Display line 3: measured flow rate in per cent of the maximum flow rate Display line 4: the character of % ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 12: Last Error

    ImpGal/s ImpGal/min To move upward in the menu, ImpGal/h use button ▲ ImpGal/day ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 13: Volume Units

    The selected fluid volume unit will appear in Basic screen. Return to Main menu: 1x ESC Return to Basic screen: ESC 2 s ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 14: Backlighting Time

    Volume flow rates in litres per second corresponding to various flow rate units Unit (litres per sec.) 1.000000E+00 l/min 1.666667E-02 2.777778E-04 Mil L/day 1.157407E+01 ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 15: Analog Output

    Flow rate for 4mA Flow rate for 20mA Output Q− .. Q+ ENTER Flow rate for 4mA +1.000000E+00 ENTER ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 16: Output For 0

    The setting procedure is essentially the same as that described for Analog output Q− ... Q+. The only value to be set here is the flow rate corresponding to output current 20mA. ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 17: Fixed Current 4-20 Ma

    Impulses for |Q| Constantly closed Q>Qlim To move upward in the menu, use button ▲ |Q|>Qlim No error ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 18: Frequency For Q

    See the procedure for setting the frequency for Q+. 5.1.4.4.3. Frequency for |Q| See the procedure for setting the frequency for Q+. ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 19: Impulses For Q

    Breaching this condition will result in an error message (E10, E11). For impulse width it holds:  ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 20 ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 21: Elis Plzeň A. S. Meter Control Instructions

    The volume per impulse parameters are selected from a decimal series; the maximum flow rate corresponds to fluid flow velocity v = 10m/s. ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 22: Impulses For Q

    Hysteresis (H) is assumed to be symmetric with respect to the flow rate limit value. Q lim ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 23: Iqi > Qlim

    The meter address can only be selected from the range of 1 to 247. Other address values will be rejected. The initial flowmeter address is 4. ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 24: Date Setting

    (Date setting). Return to Basic screen: ESC 2 s ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 25: Time Setting

    Confirm the setting by ENTER whereby the display will return to the previous menu item (Number of samples). Return to Basic screen: ESC 2 s 5.1.4.10. Suppressed flow rate Suppressed flow rate +1.000000E+00 ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 26 7,7778 3 889 10,0000 5 000 12,7778 6 389 15,5556 1000 7 778 22,2222 1200 11 111 ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 27: Zero Setting

    Return to previous menu item: ENTER Return to Main menu: 1 x ESC Return to Basic screen: ESC 2 s ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 28: Production Setting

    All outputs (analog, OUT1 and OUT2 outputs and the figure on the instantaneous flow rate line on the display) will return to the status existing prior to the commencement of the simulation mode procedure. ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 29: Remote Meter Control Using Program Floset 4.0

     Run program FLOSET 4.0, select and open the *.flo file associated with the given flowmeter type. ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 30: Program Floset 4.0 Installation And Operational Start

    ID will appear in red in the field for Meter type. Should the connection attempt fail, red label ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 31 Setting the user passwords;  Reading of the historical flow-rate data (the hourly, daily and monthly flow-rate values) ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 32: Documentation And Standards

    Communication interface RS-485 MODBUS RTU Electromagnetic flowmeters FLONEX FXx11x and FLONET FH30xx Es 90688K Configuration program FLOSET 4.0 ELIS PLZEŇ a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722 Es 90 686 K...
  • Page 33: Flonex Fxx11X And Flonet Fh30Xx Page 3 O F