GE HEALTHCARE
REV 8
SECTION 2
VISUAL INSPECTION
Inspect
2−1
Operator's Console
2−2
Collimator
Requirements are pass or fail. Service is required if any Requirement is failed. After each
Requirement is verified, place a check in its box. Record notes and comments next to
the Requirement.
Requirement
□
Switch and display markings clearly legible.
□ Covering over kVp Switch, mAs Switch,
and Displays free from cuts or punctures
□
Verify good condition of ON and OFF
symbols on console label.
□ Verify good condition of X−ray emission
symbol on console label.
□ Field Light and SID markings clearly legible.
□ Field Light switch and Field Size knobs not
loose or damaged.
□ Guard Arms perpendicular to collimator
bottom.
□ SID indicating tape attached to left side.
□ Rating Plate installed on right side of
collimator.
□ Rating Plate legible.
□ Rotational Detent knob on right side of
Collimator.
□ Must be no movement between collimator
and tube unit.
□
Verify good condition of Transverse (Lateral)
AND Longitudinal Collimator Blade symbols
on collimator labels.
□ Verify good condition of Field Light symbol
on collimator label.
□ Verify good condition of Positioning Control
Lock symbol on labels, one on EACH
collimator handle.
17
AMX−4+ FUNCTIONAL CHECK
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173224−100
Inspector's Notes