Inspect
2−3

Body

Requirement
Cassette Drawer not dented.
□ Handle attached to top of Cassette Drawer
and is free from sharp edges.
□ Drive handle is free from sharp edges.
□ No dents in left and right side covers.
□ Rubber molding attached to lower sides of
body.
□ Rubber molding attached to front bumper.
□ Exhaust Fan Port on front lower left clean
and free of obstructions.
□ Tube Support Column tight and solid in its
mounting.
Verify good condition of Type B Equipment
symbol on label on back of AMX.
□ Verify good condition of Brake Release
symbol on label next to brake release
button
.
□ Verify good condition of Hand switch
Receptacle symbol on label next to hand
switch receptacle on back of AMX.
18
AMX−4+ FUNCTIONAL CHECK
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173224−100