Technical Data - Abbott Merlin PSA EX3100 Help Manual

Pacing system analyzer
Table of Contents

Advertisement

Table 11. Parameter settings and availability
Parameter
Pulse Width
A. Refractory
V. Refractory
S1S1
Sensitivity
V. Pacing
V. Sensing
Ventricular
Support Rate
(VOO)

Technical Data

For information on the Merlin™ or Merlin™ 2 Patient Care System (PCS), refer to the Merlin PCS
Model 3650 User's manual or the Merlin 2 PCS Model MER3700 User's manual.
Mechanical Specifications (page 36)
Electrical Specifications (page 39)
LED States (page 39)
Measurement Ranges and Tolerances (page 39)
Essential Performance (page 40)
Electromagnetic Compatibility (page 40)
Battery Information (page 45)
Merlin™ PSA Accessories (page 46)
Storage and Operating Conditions (page 46)
Cleaning, Disinfecting, and Sterilization Methods (page 47)
Maintenance (page 51)
Availability
Settings
All sessions
CRT-Ds, ICDs: 0.1; 0.2; ... 1.5 ms
CRT-Ps, Pacemakers: 0.1; 0.2; ...
1.5 ms
All sessions
100; 125; ... 500 ms
All sessions
100; 125; ... 500 ms
All sessions
100; 110; ... 800 ms
All sessions
0.1; 0.2; ... 0.5; 0.75; 1.0; ... 2.0;
2.5; ... 5.0; 6.0; ... 20.0 mV
CRT-P, CRT-D
RV only; LV only; LV->RV; RV->LV;
sessions
Simultaneous
CRT-P, CRT-D
RV; LV
sessions
All sessions
Off; 30; 35; ... 95 min§
35
Nominal
0.5 ms (1.0 ms for
Burst Pacing Pulse
Width)
0.4 ms (1.0 ms for
Burst Pacing Pulse
Width)
400 ms for AAI,
AOO modes;
275 ms for ODO,
DDD, DDI, DOO,
VDD modes
250 ms
600 ms
Atrial channel:
0.5 mV
Ventricular channel:
2.0 mV
RV only
RV
Off