Select the Current of Injury button to return to the Rhythm Display.
Accessed From: Rhythm Display > Channel Control button

Waveform Control

The Waveform Control buttons on the left side of the Rhythm Display control the waveform's
appearance. To open the controls, select one of the buttons. The buttons shown depend on the
selected Source for that channel.
Figure 2. Waveform control buttons
1. Current of Injury (page 6) button (EGM only). Toggles from the filtered baseline-to-peak
view of the EGM (pictured) to the unfiltered, peak-to-peak Current of Injury view of the
EGM.
2. Configuration button. Opens the ECG Configuration (page 8) or the EGM Configuration
(page 8) window.
3. AutoGain button. Allows the programmer to automatically set an appropriate gain.
4. Plus (+) and Minus (-) buttons allow you to set the gain manually.
Accessed From: Rhythm Display > Channel Control button

Rhythm Display Setup Instructions

1.
Select the Adjust Display button to the right of the Rhythm Display.
The Adjust Display (page 8) window appears.
2.
Locate position 1.
3.
Select the Source you want to see in position 1 (ECG (page 3), EGM (page 6), Markers (page
4), or Off).
The programmer selects a default Configuration for the Source.
4.
Select the Configuration button.
If you selected ECG or EGM for the Source, the ECG Configuration (page 8) or EGM
Configuration (page 8) window opens. If you selected Markers (page 4), select the Basic or
Full button.
5.
Choose the configuration.
6.
Repeat these steps for the remaining waveforms.
7.
Select the Sweep Speed button and choose a speed.
8.
Select the ECG Filter button to enable the ECG filter
Electromagnetic interference is reduced when the ECG filter is enabled.
9.
Select the AutoGain All button to refresh the AutoGain settings.