Trouble Shooting; Trouble Shooting Icc-2000; 10.2 Trouble Shooting Iam - Pentair HAFFMANS ICC Instruction Manual

Inpack 2000 co2 calculator
Table of Contents

Advertisement

10.

TROUBLE SHOOTING

10.1 Trouble shooting ICC-2000

Trouble
Pressure drop
PET Bottle is pressed together
CO
Calculator does not react or
2
start and stops within a few
seconds
Values measured remain unstable
Display indicates "LLL" OR "HHH"

10.2 Trouble shooting IAM

Trouble
IAM tap leaks
Air volume measured is
improbably high
ICC
Reason
Upper needle leakage.
Piercing leak.
Battery voltage is less than
7 Volts.
Moisture in connector joints
has negatively affects
measurement signals.
Pressure transducer is
defective.
Temperature sensor is
defective.
The calculated CO
value is
2
out of range. The cause is
usually for out-of-range p or T
values measured.
Reason
Tap has insufficient sealing.
The lye is saturated (white)
and is no longer absorbing
CO
2.
HAFFMANS ICC
Solution
Replace O-ring (see paragraph
9.1).
Check rubber cone.
Use PET-bottle holder 005.218.
Connect the charger as
described in paragraph 9.2.
Loosen, dry and clean pressure
transducer connection cable, or
clean the temperature sensor
connection.
Check operation using
calibration procedure and
stability under atmospheric
pressure. Measured value may
not vary by more than 2 digits.
Measure the resistance between
the pins of the PT100's
connector.
Between pin 1 and 2:
pin 1 and 3:
pin 1 and 4:
pin 2 and 3:
pin 2 and 4:
pin 3 and 4:
See table 1, Appendix II
Solution
Tighten the red nut on the tap.
Renew the lye.
Page 33 of 49
110 
110 
0 
0 
110 
110 